FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 9996367 · Received April 24, 2020

Report

Report Number
3004753838-2020-038313
Event Type
Malfunction
Date Received
April 24, 2020
Date of Event
April 12, 2020
Report Date
July 2, 2020
Manufacturer
DEXCOM, INC.
Product Code
QBJ
UDI-DI
00386270000385
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459493 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9445-02 5270168 00386270000385

Patients

Seq Age Sex Outcome Treatment
1 26 YR