UPS US 200029
Report
- Report Number
- 3010266064-2020-01147
- Event Type
- Malfunction
- Date Received
- April 24, 2020
- Date of Event
- February 11, 2019
- Report Date
- April 23, 2020
- Manufacturer
- BLUE BELT TECHNOLOGIES
- Product Code
- OLO
- PMA / PMN Number
- K191223
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
H10 H3, H6: THE DEVICE WAS NOT MADE AVAILABLE TO THE DESIGNATED COMPLAINT UNIT FOR EVALUATION. THUS, VISUAL INSPECTION COULD NOT BE PERFORMED. DISCUSSION WITH THE SERVICE MANAGER FOUND THAT WHEN HE WAS SERVICING THE SYSTEM DURING PREVENTATIVE MAINTENANCE VISIT, THE UPS HAD BEEN CHARGING FOR A COUPLE OF DAYS AND THAT THE SYSTEM WAS ABLE TO BOOT. THE UPS BATTERY WAS REPLACED DURING PREVENTATIVE MAINTENANCE BECAUSE IT IS SUGGESTED THAT IT BE REPLACED EVERY 2 YEARS. AT THE TIME OF THE REPORTED EVENT ON (B)(6) 2019, THE MACHINE HAD GONE UNUSED FOR 2 MONTHS (THE PERVIOUS USE WAS (B)(6) 2018 ACCORDING TO THE FIELD REPORT). SINCE THE SYSTEM WAS UNUSED FOR SO LONG, IT NEEDED MORE THAN THE REPORTED 20 MINUTES OF CHARGING BEFORE USE.
IT WAS REPORTED THAT THERE WERE TWO SEPARATE ERRORS, BUT WE THINK BOTH OCCURRED BECAUSE THE MACHINE HAS GONE UNUSED SINCE (B)(6) 2018. THE FIRST PROBLEM RELATED TO THE INTERNAL BATTERY OF THE CPU. BECAUSE IT HAD DIED, I HAD TO RESET THE BIOS AND DATE SETTINGS BEFORE BOOTING THE SYSTEM. UPON SOLVING THAT ISSUE, I GOT ERRCODE=2000000000000 UPS BACK-UP BATTER IS NOT WORKING PROPERLY. THE LED INDICATOR ON THE FAR RIGHT, IN THE FRONT PANEL OF THE NAVIO CART WAS RED. WHEN I TALKED TO (B)(6), HE SAID I JUST HAD TO WAIT FOR IT TO GET SOME CHARGE. HOWEVER, AFTER MULTIPLE REBOOTS AND ABOUT 20 MINUTES OF CHARGING, THE LED INDICATOR REMAINED RED AND THE ERROR WOULD SHOW UP WHEN THE SYSTEM BOOTED. CASE HAD TO BE CANCELLED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 460276 | UPS US 200029 | ORTHOPEDIC STEREOTAXIC INSTRUMENT | OLO | BLUE BELT TECHNOLOGIES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |