SCREW, BONE, 4 MM OD X 5.0 (127 MM) IN LG, STRIDE
Report
- Report Number
- 3010266064-2020-00374
- Event Type
- Malfunction
- Date Received
- April 23, 2020
- Date of Event
- September 13, 2017
- Report Date
- April 23, 2020
- Manufacturer
- BLUE BELT TECHNOLOGIES
- Product Code
- OLO
- PMA / PMN Number
- K191223
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
H3, H6: THE REPORTED DEVICE, INTENDED FOR USE IN TREATMENT, WAS NOT RETURNED TO THE DESIGNATED COMPLAINT UNIT FOR INDEPENDENT EVALUATION, THUS VISUAL INSPECTION AND FUNCTIONAL TESTING COULD NOT BE PERFORMED. THE FIELD REPORT NOTED THAT SIX BONE PINS WERE BENT. THE OTHER BONE PINS WERE INVESTIGATED UNDER 17-0271-2, 17-0271-3, 17-0271-4, 17-0271-5, AND 17-0271-6. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION OR WOULD NOT BE ABLE TO PERFORM AS INTENDED. A COMPLAINT HISTORY REVIEW IDENTIFIED 22 PRIOR SIMILAR EVENTS. WE WERE NOT ABLE TO CONFIRM THERE WAS A RELATIONSHIP ESTABLISHED BETWEEN THE REPORTED EVENT AND THE DEVICE AND THE ROOT CAUSE COULD NOT BE DETERMINED. FACTORS THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT INCLUDE THE USER ALLOWING THE WEIGHT OF THE DRILL TO FALL AS IT IS ATTACHED TO THE PIN, IF THE BONE PIN IS HELD RIGIDLY IN PLACE DURING REMOVAL, OR IF THE PATIENT'S BONE IS HARDER THAN NORMAL.
IT WAS REPORTED THAT A BONE PIN OF THE NAVIO TRAY WAS FOUND BENT DURING STERILIZATION. NO PATIENT INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 455055 | SCREW, BONE, 4 MM OD X 5.0 (127 MM) IN LG, STRIDE | ORTHOPEDIC STEREOTAXIC INSTRUMENT | OLO | BLUE BELT TECHNOLOGIES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |