Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE CUSTOMERS REPORT WAS DUPLICATED AND ATTRIBUTED TO A FAULTY AUTOCAL VALVE ON THE SMART PNEUMATIC MODULE BOARD. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "RUNTIME CALIBRATION FAILURE - 1051" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED, HOWEVER IT WAS NOT A RESULT OF THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1220908-2020-01233 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "CRITICAL ALARM WARNING" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED, HOWEVER IT WAS NOT A RESULT OF THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1220908-2020-01233 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
455854 UNI-VENT, MODEL 731 VENTILATOR CBK ZOLL MEDICAL CORPORATION 799-EMVP-04 NA 00847946020620

Patients

Seq Age Sex Outcome Treatment
1 Death