FDA Adverse Event Malfunction Summary report: N

OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM

MDR report key: 9989796 · Received April 23, 2020

Report

Report Number
3004464228-2020-05864
Event Type
Malfunction
Date Received
April 23, 2020
Date of Event
April 21, 2020
Report Date
April 22, 2020
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
20385082000020
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE: MODEL: ENT450, PT-000002-GBR-ENG-MM-AW REV. 002 11/19. BLOOD GLUCOSE READINGS 4 / PAGE 55: WARNINGS: BLOOD GLUCOSE READINGS BELOW 3.9 MMOL/L MAY INDICATE HYPOGLYCAEMIA LOW BLOOD GLUCOSE). BLOOD GLUCOSE READINGS ABOVE 13.9 MMOL/L MAY INDICATE HYPERGLYCAEMIA HIGH BLOOD GLUCOSE). FOLLOW YOUR HEALTHCARE PROVIDER¿S SUGGESTIONS FOR TREATMENT.

Description of Event or Problem · 1

IT WAS REPORTED THE PINK SLIDE DID NOT MOVE FORWARD; INDICATING THE NEEDLE DID NOT DEPLOY. THE POD WAS NOT WORN AND NO ADVERSE PATIENT IMPACT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
454894 OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 18320 PD1C11041951 20385082000020

Patients

Seq Age Sex Outcome Treatment
1 37 YR