FDA Adverse Event Malfunction Summary report: N

ESTEEM +

MDR report key: 9989310 · Received April 23, 2020

Report

Report Number
9618003-2020-10368
Event Type
Malfunction
Date Received
April 23, 2020
Manufacturer
CONVATEC DOMINICAN REPUBLIC INC
Product Code
EZQ
UDI-DI
00768455188512
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY BATCH RECORD REVIEW: LOT 9M00186 WAS MANUFACTURED ON 12/04/2019, IN CONVEX 1 PC LINE, WITH A TOTAL OF (B)(4) MARKET UNITS. COMPLIANCE ENGINEER PERFORMED A BATCH RECORD REVIEW ON 08/07/2020, TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED AND NO ISSUES WERE FOUND; ALL THE COMPONENTS FOR ASSEMBLY WERE CORRECT PER BILL OF MATERIALS (BOM) AND ALL THE TOOLING INFORMATION DOCUMENTED WAS ALSO CORRECT, SYSTEM APPLICATION PRODUCT (SAP) MATERIAL ID (B)(4) AND MANUFACTURING ORDER (B)(4). THE BULK FOR THIS PRODUCT WAS MANUFACTURED IN THE LINE APS 4 AUTO UNDER LOT NUMBER 9L02944 DESCRIPTION PCH ESTINVISI DRN STD WF CLR SM CVX US , SYSTEM APPLICATION PRODUCT (SAP) MATERIAL 1701428 , QUANTITY 8,800 MANUFACTURING DATE 11/27/2019, ORDER NUMBER (B)(4). THE BATCH RECORD REVIEW SUPPORTS THAT THERE WERE NO DISCREPANCIES RELATED TO THE ISSUE REPORTED. PHOTOGRAPH, VIDEO AND/OR PHYSICAL SAMPLE EVALUATION: NO PHOTOGRAPH ASSOCIATED WITH THIS CASE WERE RECEIVED. NO UNUSED RETURN SAMPLE HAS BEEN RECEIVED FOR THIS COMPLAINT. CONCLUSION SUMMARY OF THE RELATED EVENT: BASED IN THE INVESTIGATION FINDINGS, THE ROOT CAUSE IDENTIFIED FOR THE ISSUE ¿WAFER OFF CENTER IN THE POUCH¿, REPORTED UNDER FAILURE MODE ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR¿ IS ATTRIBUTED TO: 1) MACHINE: THE INVESTIGATIVE PROCESS CONCLUDED THAT ALL THE MACHINERY/ TOOLING ITEMS USED IN THE MANUFACTURING PROCESS COMPLIED WHEN COMPARED AGAINST DRAWING AND PROCESS INSTRUCTION (PI) SPECIFICATIONS; HOWEVER, AS PART OF THIS INVESTIGATION ON 10/MAY/2019, MAINTENANCE AND FACILITIES MANAGER AND SENIOR PROCESS ENGINEER PERFORMED A DEEP ASSESSMENT TO THE SUB-ASSEMBLY STATION OF THE CONVEX 1PC MACHINE IDENTIFYING THE CONDITIONS CAUSING THE WAFER OFF CENTER AND THE POSSIBLE SOLUTIONS TO CORRECT THE FAILURE. AS CONCLUSION OF THE ASSESSMENT, THE CHAIN THAT COORDINATES FLANGE LOADING STATION WITH ADHESIVE LOADING STATION NEEDED LUBRICATION AND THE CHAIN AND OTHER COMPONENTS OF THE MECHANISM NEEDED A CORRECTIVE MAINTENANCE. IN ORDER TO MAINTAIN THIS CONDITION IN THE LONG TERM, A DEEPER INTERVENTION OF THE MECHANISM IS REQUIRED. FURTHERMORE, REGARDING THE ROTARY TABLE STATION OF THE CONVEX 1-PIECE (PC) MACHINE, IT WAS IDENTIFIED THAT DUE TO THE DEMANDS OF THE PROCESS, THE TOOLING USED TO MANUFACTURE THE URODRESS PRODUCT FAMILY REQUIRE A REDESIGN MODIFICATION TO REDUCE THE VARIABILITY WITHIN THE MANUFACTURING PROCESS. AS A CONTRIBUTOR FACTORS THE FOLLOWING OPPORTUNITIES FOR IMPROVEMENTS WERE IDENTIFIED: 2) PROCESS / METHOD INVESTIGATION: A) REFERENCE IN THE PROCEDURE INSTRUCTION (PI), THE APPLICABLE QUALITY CONTROL (QC) TOOLING USED FOR PRODUCT¿S QUALITY INSPECTION PURPOSE OF POUCH URO 177A PRODUCTS FAMILY IN ORDER TO ADDRESS MANUFACTURING PERSONNEL THE USE OF THEM. B) IDENTIFICATION OF THE DEFECT DURING THE MANUFACTURING PROCESS, IT SHOULD BE CONSIDERED THE IMPLEMENTATION OF A JOB AID FOR PRODUCT CENTRALIZATION OUT OF SPECIFICATION. 3) PROCESS / METHOD / MEASUREMENT INVESTIGATION A) INTRODUCE QUALITY CONTROL (QC) TOOLING USED FOR PRODUCT¿S QUALITY INSPECTION PURPOSE IN THE CALIBRATION PROGRAM TO GUARANTEE MEASUREMENT ASSURANCE. 4) MANPOWER: A) THERE ARE NOT MANPOWER CAUSES ASSOCIATED WITH THE ROOT CAUSE; EVEN THOUGH, OPPORTUNITY TO IMPROVE THE LOADING OF THE ADHESIVE DISC DURING THE MANUFACTURING PROCESS WAS IDENTIFIED BY IMPLEMENTING A STANDARD WORK INSTRUCTION FOR CONVEX 1-PIECE (PC) TO INDICATE MANUFACTURING OPERATORS HOW TO PROPERLY LOAD THE ADHESIVE DISC. NO ISSUES WERE IDENTIFIED FOR MATERIAL AND ENVIRONMENT INVESTIGATIONS. A CORRECTIVE ACTIONS/PREVENTIVE ACTIONS (CAPA) PLAN WAS GENERATED FOR THE MITIGATE THE ROOT CAUSES IDENTIFIED. THE INVESTIGATION ASSOCIATED WITH RELATED EVENT HAS BEEN APPROVED AND IS COMPLETE. NO ADDITIONAL ACTION IS REQUIRED, AND THIS COMPLAINT WILL BE CLOSED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER: REPORTING SITE: 1049092. MANUFACTURING SITE: 9618003.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Additional Manufacturer Narrative · 1

DEVICE 2 OF 4. BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THE END USER REPORTED THAT 4 OUT OF 10 APPLIANCES HAD AN OFF CENTERED WAFER STOMA OPENING. THE AFFECTED WAFERS HAD A DECREASED WEAR TIME OF 1 DAY. THERE WAS NO HARM REPORTED BY THE END USER. NO PHOTO IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
454784 ESTEEM + POUCH, COLOSTOMY EZQ CONVATEC DOMINICAN REPUBLIC INC 416742 9M00186 00768455188512

Patients

Seq Age Sex Outcome Treatment
1 Female