FDA Adverse Event Malfunction Summary report: N

INFINITY DOCKING STATION

MDR report key: 998175 · Received February 8, 2008

Report

Report Number
1220063-2008-00003
Event Type
Malfunction
Date Received
February 8, 2008
Date of Event
December 20, 2007
Report Date
February 7, 2008
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
MHX
Removal / Correction Number
2517967-11/16/06-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUR EVALUATION CONFIRMED THAT THE TOP COVER OF THE DEVICES ACTUALLY SEPARATED FROM THE BASE AS REPORTED. WE ALSO FOUND THAT ONE OF THE UNITS WAS MODIFIED (TOP COVER GLUED TO BASE) BY THE CUSTOMER. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICES IDENTIFIED IN THIS INCIDENT FALL IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION.

Description of Event or Problem · 1

DRAEGER MEDICAL SYSTEMS, INC. HAS RECEIVED INFORMATION FROM A CUSTOMER WHO HAS REPORTED THAT TWO OF OUR INFINITY DOCKING STATIONS (IDS) UNITS WERE FOUND TO HAVE SEPARATED (TOP COVER FROM THE BASE COMPONENT). NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITY DOCKING STATION PATIENT MONITOR DOCKING STATION MHX DRAEGER MEDICAL SYSTEMS, INC. 5206110 NA

Patients

Seq Age Sex Outcome Treatment
1 NI