FDA Adverse Event
Malfunction
Summary report: N
INFINITY DOCKING STATION
MDR report key: 998175
·
Received February 8, 2008
Report
- Report Number
- 1220063-2008-00003
- Event Type
- Malfunction
- Date Received
- February 8, 2008
- Date of Event
- December 20, 2007
- Report Date
- February 7, 2008
- Manufacturer
- DRAEGER MEDICAL SYSTEMS, INC.
- Product Code
- MHX
- Removal / Correction Number
- 2517967-11/16/06-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
OUR EVALUATION CONFIRMED THAT THE TOP COVER OF THE DEVICES ACTUALLY SEPARATED FROM THE BASE AS REPORTED. WE ALSO FOUND THAT ONE OF THE UNITS WAS MODIFIED (TOP COVER GLUED TO BASE) BY THE CUSTOMER. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICES IDENTIFIED IN THIS INCIDENT FALL IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION.
Description of Event or Problem · 1
DRAEGER MEDICAL SYSTEMS, INC. HAS RECEIVED INFORMATION FROM A CUSTOMER WHO HAS REPORTED THAT TWO OF OUR INFINITY DOCKING STATIONS (IDS) UNITS WERE FOUND TO HAVE SEPARATED (TOP COVER FROM THE BASE COMPONENT). NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITY DOCKING STATION | PATIENT MONITOR DOCKING STATION | MHX | DRAEGER MEDICAL SYSTEMS, INC. | 5206110 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |