UNKNOWN_MEDICAL - REDMOND_PRODUCT
Report
- Report Number
- 0003015876-2020-00532
- Event Type
- Injury
- Date Received
- April 20, 2020
- Date of Event
- March 1, 2020
- Report Date
- April 21, 2020
- Manufacturer
- PHYSIO-CONTROL, INC. - 3015876
- Product Code
- DRM
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SE
- Reporter Occupation
- OTHER
Narratives
SECTION D2 PRODUCT CODE # OF INITIAL MEDWATCH REPORT INDICATES: MKJ. SECTION D2 PRODUCT CODE # OF THIS MEDWATCH REPORT INDICATES: DRM.
PHYSIO-CONTROL WAS MADE AWARE OF THE ARTICLE "COMPARTMENT SYNDROME OF THE HAND AS A COMPLICATION OF PROLONGED MECHANICAL CARDIOPULMONARY RESUSCITATION" PUBLISHED IN ANAESTHESIA REPORTS 2020-8, THAT MENTIONED A PATIENT EVENT WHERE THE DEVICE USE HAD POTENTIALLY CAUSED COMPARTMENT SYNDROME IN A 45-YEAR-OLD MALE PATIENT DURING MECHANICAL CARDIOPULMONARY RESUSCITATION. CARDIOPULMONARY RESUSCITATION ALONG WITH AGGRESSIVE REWARMING ACHIEVED RETURN OF SPONTANEOUS CIRCULATION. DESPITE PRESERVATION OF NEUROLOGICAL FUNCTION THE PATIENT DIED OF COMPLICATIONS FROM THE CARDIAC ARREST AFTER AN EXTENDED INTENSIVE CARE UNIT STAY.
PHYSIO-CONTROL PERFORMED A CLINICAL REVIEW OF THE REPORTED EVENT AND DETERMINED THAT USER ERROR CONTRIBUTED TO THE SERIOUS INJURY OF THE PATIENT. THE PATIENT HAD HIS ARM STRAPPED TO THE DEVICE WITH THE PATIENT STRAPS DURING THE PROLONGED RESUSCITATION. THIS RESULTED IN COMPARTMENT SYNDROME TO THE PATIENT'S LEFT HAND. USER ERROR.
PHYSIO-CONTROL WAS MADE AWARE OF THE ARTICLE "COMPARTMENT SYNDROME OF THE HAND AS A COMPLICATION OF PROLONGED MECHANICAL CARDIOPULMONARY RESUSCITATION" PUBLISHED IN ANAESTHESIA REPORTS 2020-8, THAT MENTIONED A PATIENT EVENT WHERE THE DEVICE USE HAD POTENTIALLY CAUSED COMPARTMENT SYNDROME IN A (B)(6)-YEAR-OLD MALE PATIENT DURING MECHANICAL CARDIOPULMONARY RESUSCITATION. CARDIOPULMONARY RESUSCITATION ALONG WITH AGGRESSIVE REWARMING ACHIEVED RETURN OF SPONTANEOUS CIRCULATION. DESPITE PRESERVATION OF NEUROLOGICAL FUNCTION THE PATIENT DIED OF COMPLICATIONS FROM THE CARDIAC ARREST AFTER AN EXTENDED INTENSIVE CARE UNIT STAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 440460 | UNKNOWN_MEDICAL - REDMOND_PRODUCT | COMPRESSOR, CARDIAC, EXTERNAL | DRM | PHYSIO-CONTROL, INC. - 3015876 | LUCAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Other |