FDA Adverse Event Malfunction Summary report: N

PRIMARY PLUM SET, 2 CLAVE MULTIPORT CONNECTOR, BALL CHECK VALVE IN SIGHT CHAMBER

MDR report key: 9973742 · Received April 17, 2020

Report

Report Number
9615050-2020-00056
Event Type
Malfunction
Date Received
April 17, 2020
Date of Event
March 25, 2020
Report Date
March 31, 2020
Manufacturer
ICU MEDICAL COSTA RICA LTD.
Product Code
FPA
PMA / PMN Number
K982159
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H10 - THE COMPLAINT OF LEAKAGE CAN BE CONFIRMED ON THE RETURNED ONE USED 121933190 PRIMARY PLUM SET. THE SECONDARY PORT WAS BENT AND THERE WAS A BREAK OBSERVED ON THE CASSETTE AS RECEIVED. THERE WAS PLASTIC DEFORMATION OBSERVED ON THE SECONDARY PORT. THE SET LEAKED FROM THE BREAK OBSERVED. THE PROBABLE CAUSE OF THE DAMAGE OBSERVED IS UNKNOWN HOWEVER, IT WOULD HAVE HAPPENED OUTSIDE ICU MEDICAL. A DEVICE HISTORY REVIEW FOR LOT# 4040976 AND RELEVANT COMMODITIES WERE REVIEWED, AND NO NON-CONFORMANCES WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. ADDITIONAL INFORMATION CAN BE FOUND IN SECTION G1.

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO RETURN FOR EVALUATION. IT HAS NOT BEEN RECEIVED.

Description of Event or Problem · 1

THE EVENT INVOLVED A PRIMARY PLUMSET THAT DURING CHEMOTHERAPY TREATMENT A LEAK WAS OBSERVED. THE PATIENT WAS RECEIVING THE 5 CYCLE OF POLY-CHEMOTHERAPY APPLICATION, DAY 15 OF THE FLOT PROTOCOL (DOCETAXEL + CALCIUM FOLINATE IN OXALIPLATIN, 5 FLUOROURACIL IN INFUSER). AT 14 O¿CLOCK, WHEN THE MIXTURE OF OXALIPLATIN IN CALCIUM FOLINATE WAS ALREADY INFUSING, A LEAK WAS OBSERVED FROM THE SET LINE B, WHERE THE CALCIUM FOLINATE WAS CONNECTED. THE INFUSION WAS STOPPED, THE SET WAS CHECKED, AND IT WAS NOTED THAT THE SET WAS BROKEN IN THE LINE B, CONNECTOR. THE TUBING SET WAS REPLACED. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO HARMED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435887 PRIMARY PLUM SET, 2 CLAVE MULTIPORT CONNECTOR, BALL CHECK VALVE IN SIGHT CHAMBER SET, ADMINISTRATION, INTRAVASCULAR FPA ICU MEDICAL COSTA RICA LTD. 4040976

Patients

Seq Age Sex Outcome Treatment
1 5 FLUOROURACIL, MFR UNK| CALCIUM FOLINATE, MFR UNK| DOCETAXEL, MFR UNK| OXALIPLATIN, MFR UNK