VENTANA HE 600 SYSTEM

MDR report key: 9972220 · Received April 17, 2020

Report

Report Number
2028492-2020-00005
Event Type
Malfunction
Date Received
April 17, 2020
Date of Event
March 24, 2020
Report Date
December 22, 2020
Manufacturer
VENTANA MEDICAL SYSTEMS INC
Product Code
KPA
PMA / PMN Number
510K EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THROUGH THE CAPA INVESTIGATION, IMPROVEMENTS HAVE BEEN MADE TO THE VENTANA DESIGN PROCESS, INCLUDING DESIGN READINESS AND MATERIAL COMPATIBILITY WORK INSTRUCTIONS AND CHECKLISTS. WE ARE CURRENTLY MONITORING THE PROGRESS OF THE UPDATES MADE TO THE VENTANA DESIGN PROCESSES TO ENSURE EFFECTIVE IMPLEMENTATION. (B)(4).

Additional Manufacturer Narrative · 1

THE ROCHE FIELD ENGINEER REPLACED THE AFFECTED PARTS. THE INSTRUMENT HAS BEEN FIXED AND HANDED BACK TO THE CUSTOMER FOR OPERATION. THE DEVICE'S UDI NUMBER IS: (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE SITE SMELLED A BURNING ODOR COMING FROM THE INSTRUMENT AND THE INSTRUMENT DISPLAYED AN ERROR MESSAGE. THE SITE EVACUATED THE BUILDING AS A PRECAUTION. A ROCHE FIELD SERVICE ENGINEER (FSE) WENT ON SITE AND FOUND A PIN-HOLE ON THE T-FITTING THAT CAUSED A SLOW LEAK OF THE HE 600 CLEANING SOLUTION TO THE COMPRESSOR BAY. LIQUID DRIPPED ON THE VENT FAN CONNECTOR AND CREATED AN ELECTRICAL SHORT THAT STOPPED THE FAN. WITHOUT THE FAN WORKING, THE COMPRESSOR OVERHEATED AND THE ELEVATED TEMPERATURE MELTED THE PLASTIC FAN CONNECTOR THAT PRODUCED A BURNING SMELL AND GENERATED THE ERROR MESSAGE. THE ENGINEER INDICATED THERE WAS NO EVIDENCE OF FIRE FOUND AND THERE WAS NO HARM TO ANY USERS ON SITE. THE INSTRUMENT HAS BEEN FIXED AND TESTED BY THE FSE, AND HANDED BACK OVER TO THE CUSTOMER FOR OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435724 VENTANA HE 600 SYSTEM SLIDE STAINER, AUTOMATED KPA VENTANA MEDICAL SYSTEMS INC NA NA