FDA Adverse Event Malfunction Summary report: N

TRUETRACK

MDR report key: 9967763 · Received April 16, 2020

Report

Report Number
1000113657-2020-00240
Event Type
Malfunction
Date Received
April 16, 2020
Date of Event
March 23, 2020
Report Date
April 16, 2020
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292008642
PMA / PMN Number
K032657
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER WAS RETURNED FOR EVALUATION. DEFECT WAS DETECTED: BROKEN PIECE OF METER CASE (STRIP CONNECTOR). TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: RC-16: BENT PINS ON THE STRIP CONNECTOR CAUSED BY THE USER NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 241, 221, 252, 138 AND 245 MG/DL; THE RESULT OF 241 MG/DL WAS NON-FASTING. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT IS 120 MG/DL, EXPECTED POST-LUNCH NON-FASTING BLOOD GLUCOSE TEST RESULT IS 125 MG/DL AND EXPECTED PRE-DINNER FASTING BLOOD GLUCOSE TEST RESULT IS 149 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 03/16/2022 AND OPEN VIAL DATE IS (B)(6) 2020. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (TIME NOT SET): (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
432048 TRUETRACK SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, CVS 100CTTT RW5372S 00021292008642

Patients

Seq Age Sex Outcome Treatment
1