LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS
Report
- Report Number
- 2939274-2020-01862
- Event Type
- Malfunction
- Date Received
- April 16, 2020
- Report Date
- March 23, 2020
- Manufacturer
- WRIGHTS LANE SYNTHES USA PRODUCTS LLC
- Product Code
- HTJ
- UDI-DI
- 10886982196255
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PART NUMBER: 357.402. SYNTHES LOT NUMBER: 5760030. SUPPLIER LOT NUMBER: N/A. RELEASE TO WAREHOUSE DATE: APRIL 10, 2008. EXPIRATION DATE: N/A. MANUFACTURED BY SYNTHES BRANDYWINE. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS (P/N: 357.402, LOT NUMBER: 5760030) WAS RECEIVED AT US CUSTOMER QUALITY (CQ). UPON VISUAL INSPECTION, THE BALL AND COMPRESSION SPRING COMPONENTS ARE MISSING. THE SHAFT IS ALSO BENT IN THE MIDDLE. DEVICE FAILURE/DEFECT IDENTIFIED? YES. DIMENSIONAL INSPECTION: DRAWING. SPECIFIED DIMENSIONS: SHAFT DIAMETER = 2.5 +0/-0.1MM. MEASURED DIMENSIONS: SHAFT DIAMETER = 2.45MM, CONFORMING. DEVICE USED - CALIPER CA802. DIMENSIONAL ANALYSIS ON THE HOLE FOR THE MISSING BALL AND COMPRESSION SPRING COMPONENTS COULD NOT BE PERFORMED DUE TO DEFORMATION FROM THE STAKING PROCESS THAT HAPPENS DURING FINAL ASSEMBLY. DOCUMENT/SPECIFICATION REVIEW: CURRENT AND MANUFACTURED WERE REVIEWED. CURRENT AND MANUFACTURED WERE ALSO REVIEWED. NO DESIGN ISSUES OR DISCREPANCIES WERE IDENTIFIED. COMPLAINT CONFIRMED? YES. CONCLUSION: THE OVERALL COMPLAINT WAS CONFIRMED FOR THE LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS (P/N: 357.402, LOT NUMBER: 5760030) AS THE BALL AND COMPRESSION SPRING COMPONENTS ARE MISSING AND THE SHAFT IS BENT IN THE MIDDLE. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED BASED ON THE PROVIDED INFORMATION. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. NO DESIGN ISSUES WERE OBSERVED DURING THE DOCUMENT/SPECIFICATION REVIEW. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT ON AN UNKNOWN DATE, A LOCKING BOLT MEASURING DEVICE WAS FOUND TO BE BROKEN DURING REVERSE LOGISTICS AUDIT OF THE RETURNED DEVICE AT MILLSTONE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431539 | LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS | GAUGE,DEPTH | HTJ | WRIGHTS LANE SYNTHES USA PRODUCTS LLC | 357.402 | 5760030 | 10886982196255 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |