FDA Adverse Event Malfunction Summary report: N

CONNECTA WHT 360DEG TB 25CM

MDR report key: 9963052 · Received April 15, 2020

Report

Report Number
9610847-2020-00135
Event Type
Malfunction
Date Received
April 15, 2020
Date of Event
March 17, 2020
Report Date
July 31, 2020
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMG
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVALUATION?: YES. D.10. RETURNED TO MANUFACTURER ON: (B)(6)2020. H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED LOT NUMBER 9193077. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT AND ALL INSPECTIONS WERE FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS INCIDENT, ONE PHYSICAL SAMPLE WAS RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE SAMPLE, THE ADAPTER COMPONENT WAS OBSERVED SEPARATED FROM THE TUBING. IT HAS BEEN DETERMINED THAT THIS SEPARATION WAS A RESULT OF PARTIAL ADHESIVE SOLVENT BETWEEN THE TWO COMPONENTS, WHICH WOULD ALSO RESULT IN LEAKAGE. A QUALITY ALERT HAS BEEN RAISED IN RESPONSE TO THIS INCIDENT.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED LOT NUMBER 9193077. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF CONNECTAS WHT 360DEG TB 25CM EXPERIENCED TUBING SEPARATION FROM ADAPTER/LUER CONNECTION. PRODUCT DEFECT WAS NOTED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: EACH TIME YOU FLUSH THE ACCESS FROM THE EXTENSION CORD, IT DISCONNECTS FROM THE PROXIMAL CONNECTION TO THE VENOUS ACCESS. LEAKAGE OF THE INFUSION WITH DISPERSION AND CONTAMINATION OF THE DRUG.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF CONNECTAS WHT 360DEG TB 25CM EXPERIENCED TUBING SEPARATION FROM ADAPTER/LUER CONNECTION. PRODUCT DEFECT WAS NOTED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: EACH TIME YOU FLUSH THE ACCESS FROM THE EXTENSION CORD, IT DISCONNECTS FROM THE PROXIMAL CONNECTION TO THE VENOUS ACCESS. LEAKAGE OF THE INFUSION WITH DISPERSION AND CONTAMINATION OF THE DRUG.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF CONNECTAS WHT 360DEG TB 25CM EXPERIENCED TUBING SEPARATION FROM ADAPTER/LUER CONNECTION. PRODUCT DEFECT WAS NOTED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: EACH TIME YOU FLUSH THE ACCESS FROM THE EXTENSION CORD, IT DISCONNECTS FROM THE PROXIMAL CONNECTION TO THE VENOUS ACCESS. LEAKAGE OF THE INFUSION WITH DISPERSION AND CONTAMINATION OF THE DRUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
429575 CONNECTA WHT 360DEG TB 25CM STOPCOCK FMG BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 9193077

Patients

Seq Age Sex Outcome Treatment
1 Other