FDA Adverse Event Malfunction Summary report: N

IMPL TWIST MP-1 5.0 MM 10 MM

MDR report key: 9962271 · Received April 15, 2020

Report

Report Number
0002023141-2020-00770
Event Type
Malfunction
Date Received
April 15, 2020
Date of Event
February 29, 2020
Report Date
June 18, 2020
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K013494
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

IMPLANT WAS RETURNED FOR INVESTIGATION. DEVICE HAD NO DRIED BLOOD OR SIGNS OF USE AND PASSED FUNCTIONAL TESTING AS THE MOUNT WAS EASILY REMOVED. NO PRE-EXISTING CONDITIONS WERE NOTED ON THE PER AND THE PATIENT HAD UNKNOWN BONE DENSITY. THE REPORTED DEVICE WAS INTENDED FOR USE ON TOOTH #19 (UNIVERSAL) AND WERE USED FOR A SINGLE DAY/PROCEDURE. THE CUSTOMER DID NOT PROVIDE PICTURES OR ADDITIONAL EVIDENCE. APPROPRIATE DOCUMENTATION WAS REVIEWED. A DHR REVIEWS WERE COMPLETED FOR THE SUBJECT LOT NUMBER. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHRS. LOTS WERE INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. A COMPLAINT HISTORY REVIEWS WERE PERFORMED FOR THE REPORTED LOT NUMBER FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID NOT OCCUR AND THE REPORTED EVENT WAS UNCONFIRMED AS THE DEVICE PASSED FUNCTIONAL TESTING AS THE MOUNT WAS EASILY REMOVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. PATIENT SEX UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED. EXPIRATION DATE AND UDI NUMBER UNKNOWN / NOT PROVIDED. PMA/510K: K962106.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTS BODY CANNOT BE REMOVED FROM THE DELIVERY MOUNTS. THE IMPLANTS WERE REMOVED AND ANOTHER IMPLANTS WERE PLACED. TOOTH LOCATION #19.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
424642 IMPL TWIST MP-1 5.0 MM 10 MM DENTAL IMPLANT DZE ZIMMER DENTAL 63101822

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention