T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2020-37063
- Event Type
- Malfunction
- Date Received
- April 14, 2020
- Date of Event
- March 23, 2020
- Report Date
- April 14, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OZO
- UDI-DI
- 00853052007264
- PMA / PMN Number
- P180008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE T:SLIM X2 BASAL IQ USER GUIDE STATES: DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT MULTIPLE OCCLUSION ALARMS OCCURRED. CUSTOMER REPORTED USING AMDELOG INSULIN. AMDELOG IS OFF LABEL PER THE USER GUIDE. CUSTOMER EITHER RESUMED INSULIN OR CHANGED PUMP SUPPLIES TO ADDRESS THE OCCLUSION ALARMS AND SUCCESSFULLY RESUMED INSULIN. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 250-460 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421233 | T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | TANDEM DIABETES CARE | 1000096 | 00853052007264 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |