FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 9956509 · Received April 14, 2020

Report

Report Number
2032227-2020-102748
Event Type
Malfunction
Date Received
April 14, 2020
Date of Event
December 8, 2019
Report Date
June 15, 2020
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
00763000090203
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE PASSED DISPLACEMENT TEST. DEVICE RECEIVED WITH SAME AUDIO TONE WHEN SWITCHING FROM VOLUME 5 TO VOLUME 1, PUMP ERROR 63 (VARIABLE 3) WAS PRESENT IN THE DEVICE TRACE DOWNLOAD AND PUMP ERROR 63 (VARIABLE 3) ALARMED DURING SELFTEST DUE TO BROKEN TRACE AT U1 PIN 6 (SDA) ON KEYPAD ASSEMBLY.

Additional Manufacturer Narrative · 1

DEVICE PASSED DISPLACEMENT TEST. DEVICE RECEIVED WITH SAME AUDIO TONE WHEN SWITCHING FROM VOLUME 5 TO VOLUME 1, PUMP ERROR 63 (VARIABLE 3) WAS PRESENT IN THE DEVICE TRACE DOWNLOAD AND PUMP ERROR 63 (VARIABLE 3) ALARMED DURING SELFTEST DUE TO BROKEN TRACE AT U1 PIN 6 (SDA) ON KEYPAD ASSEMBLY.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD AN AUDIO ANOMALY. THE DEVICE FAILED THE AUDIO TEST. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
423090 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG2KVJJ 00763000090203

Patients

Seq Age Sex Outcome Treatment
1 49 YR