PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 2032227-2020-102748
- Event Type
- Malfunction
- Date Received
- April 14, 2020
- Date of Event
- December 8, 2019
- Report Date
- June 15, 2020
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 00763000090203
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
DEVICE PASSED DISPLACEMENT TEST. DEVICE RECEIVED WITH SAME AUDIO TONE WHEN SWITCHING FROM VOLUME 5 TO VOLUME 1, PUMP ERROR 63 (VARIABLE 3) WAS PRESENT IN THE DEVICE TRACE DOWNLOAD AND PUMP ERROR 63 (VARIABLE 3) ALARMED DURING SELFTEST DUE TO BROKEN TRACE AT U1 PIN 6 (SDA) ON KEYPAD ASSEMBLY.
DEVICE PASSED DISPLACEMENT TEST. DEVICE RECEIVED WITH SAME AUDIO TONE WHEN SWITCHING FROM VOLUME 5 TO VOLUME 1, PUMP ERROR 63 (VARIABLE 3) WAS PRESENT IN THE DEVICE TRACE DOWNLOAD AND PUMP ERROR 63 (VARIABLE 3) ALARMED DURING SELFTEST DUE TO BROKEN TRACE AT U1 PIN 6 (SDA) ON KEYPAD ASSEMBLY.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD AN AUDIO ANOMALY. THE DEVICE FAILED THE AUDIO TEST. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 423090 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG2KVJJ | 00763000090203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |