FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 9955338 · Received April 13, 2020

Report

Report Number
9617229-2020-06700
Event Type
Injury
Date Received
April 13, 2020
Report Date
June 17, 2020
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON MAY 06, 2020 WITH LOT NUMBER 2895523. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED; CREASES FOLD, DEFORMATION, YELLOW PARTICLES ON THE SHELL AND THE WEIGHT THE DEVICE WITHIN SPECIFICATION. BUBBLES IN THE GEL OBSERVED AFTER AUTOCLAVE CYCLE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS.

Description of Event or Problem · 0

PHYSICIAN REPORTED " CAPSULAR CONTRACTURE - BAKER GRADE III". DEVICE HAS BEEN EXPLANTED. THIS RELATES TO THE LEFT SIDE.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN INITIATED. IF ANY NEW, CHANGED OR CORRECTED INFORMATION IS NOTED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED. THE EVENT OF CAPSULAR CONTRACTURE BAKER GRADE III IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE III.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
417265 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2895523

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention