FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 9950725 · Received April 10, 2020

Report

Report Number
3004753838-2020-034743
Event Type
Malfunction
Date Received
April 10, 2020
Date of Event
March 6, 2020
Report Date
June 8, 2020
Manufacturer
DEXCOM, INC.
Product Code
PQF
UDI-DI
10386270000221
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. A REVIEW OF THE SHARE LOG WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411603 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR PQF DEXCOM, INC. 9438-06 7259842 10386270000221

Patients

Seq Age Sex Outcome Treatment
1 30 YR