FDA Adverse Event Malfunction Summary report: N

BD NEEDLE 18GA 1IN BLUNT FILL SLA

MDR report key: 9935979 · Received April 8, 2020

Report

Report Number
3003916417-2020-00114
Event Type
Malfunction
Date Received
April 8, 2020
Date of Event
March 17, 2020
Report Date
April 16, 2020
Manufacturer
BECTON DICKINSON IND. CIRURGICAS LTDA
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BATCH HISTORY ANALYSIS (DHR), MAINTENANCE RECORDS AND QUALITY NOTIFICATIONS WERE VERIFIED AND A MAINTENANCE RECORD WAS FOUND ON THE NEEDLE WRAPPING MACHINE RELATED TO THE DAMAGED PROTECTIVE DEFECT CLAIMED BY THE CUSTOMER FOR THIS BATCH. WE RECEIVED THE SAMPLE SENT FOR EVALUATION, SO IT WAS POSSIBLE TO CONDUCT AN INVESTIGATION, CONFIRM A CLAIM FOR THE BROKEN PROTECTIVE DEFECT AND DETERMINE THE ROOT CAUSE. THE INCIDENT OCCURRED DUE TO A PROTECTOR ENTANGLEMENT IN THE ASSEMBLY LINE BECOMING MISALIGNED CAUSING THE INCIDENT. H3 OTHER TEXT : SEE H.10

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE THE NEEDLE TIP WAS DISCOVERED TO BE DEFECTIVE WITH A BD NEEDLE (B)(4) GA (B)(4) IN BLUNT FILL SLA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: NEEDLE TIP IS IRREGULAR.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE #: UNKNOWN. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE THE NEEDLE TIP WAS DISCOVERED TO BE DEFECTIVE WITH A BD NEEDLE 18GA 1IN BLUNT FILL SLA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: NEEDLE TIP IS IRREGULAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
400265 BD NEEDLE 18GA 1IN BLUNT FILL SLA HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON IND. CIRURGICAS LTDA 9197336

Patients

Seq Age Sex Outcome Treatment
1 Other