FDA Adverse Event Injury Summary report: N

T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY

MDR report key: 9934192 · Received April 8, 2020

Report

Report Number
3013756811-2020-33642
Event Type
Injury
Date Received
April 8, 2020
Date of Event
December 19, 2019
Report Date
April 7, 2020
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00853052007264
PMA / PMN Number
DEN180058
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

TANDEM QUALITY ENGINEER EVALUATED PUMP DATA AND CONCLUDED THE FOLLOWING: LOGS WERE REVIEWED FROM 2/15/2020 TO 3/19/2020. A TOTAL OF 5 LOW EVENTS WERE DETECTED WITHIN THE SPECIFIED DATE AND TIME RANGE. A LOW EVENT IS DEFINED AS A GLUCOSE LESS THAN 54 MG/DL IN A 2-HOUR PERIOD. THE RESULTS OF THE INVESTIGATION FOR THIS EVENT IS INCLUDED BELOW. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 50 MG/DL WERE RECORDED AT 01:17:35 AM ON 2/16/2020. A USER INITIATED TUBING FILL OF SIZE 10.55 UNITS WAS OBSERVED AT 10:39:28 PM ON 2/15/2020. IF THE USER DID NOT DISCONNECT DURING THE ¿FILL TUBING¿ STEP OF THE CARTRIDGE CHANGE SEQUENCE, INSULIN WOULD BE DELIVERED, AND THIS COULD CAUSE BG LEVELS TO DROP. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 53 MG/DL WERE RECORDED AT 07:47:37 AM TO 10:27:38 AM ON 2/18/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 55 WAS ENUNCIATED AT 07:42:37 AM ON 2/18/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF 50 TO 52 MG/DL WERE RECORDED AT 5:27:48 PM TO 5:42:48 PM ON 2/23/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 55 WAS ENUNCIATED AT 5:22:48 PM ON 2/23/2020. THE USER MADE THE FOLLOWING BOLUS REQUEST(S) WHILE HAVING INSULIN ON BOARD (IOB): TIME: 12:23:16 PM DATE: 2/23/2020 IOB: 1.18 REQUESTING A BOLUS WITH IOB MAY LEAD TO A LOW GLUCOSE EVENT. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 51 MG/DL WERE RECORDED AT 9:32:49 PM TO 9:47:49 PM ON 2/23/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 51 WAS ENUNCIATED AT 9:32:49 PM ON 2/23/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF 42 TO 47 MG/DL WERE RECORDED AT 10:02:49 PM ON 2/23/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 51 WAS ENUNCIATED AT 9:32:49 PM ON 2/23/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. THERE IS NO EVIDENCE THAT THE PUMP EXPERIENCED A MALFUNCTION OR FAILURE.

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED A LOW BLOOD GLUCOSE (BG) LEVEL BETWEEN 40-90 MG/DL. BG CAUSE WAS NOT KNOWN. CUSTOMER DRANK A JUICE TO ADDRESS BG. RECOMMENDATION WAS MADE TO CONSULT WITH A HEALTHCARE PROVIDER TO DISCUSS DIABETES MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403417 T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1000096 00853052007264

Patients

Seq Age Sex Outcome Treatment
1 47 YR Other