FDA Adverse Event Malfunction Summary report: N

CAPTURA SERRATED FORCEPS W/O SPIKE

MDR report key: 9933381 · Received April 7, 2020

Report

Report Number
1037905-2020-00168
Event Type
Malfunction
Date Received
April 7, 2020
Date of Event
March 6, 2020
Report Date
March 9, 2020
Manufacturer
COOK ENDOSCOPY
Product Code
PTS
UDI-DI
00827002561250
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

OCCUPATION: UNKNOWN. INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT CONFIRMED THE COMPLAINT AS DESCRIBED BY THE USER. THE DEVICE WAS VISUALLY EXAMINED, AND THE CUPS APPEAR TO BE POTENTIALLY MISALIGNED. DURING A FUNCTIONAL TEST, THE HANDLE WAS MANIPULATED, AND THE CUPS WOULD OPEN AS INTENDED BUT DID NOT CLOSE AS INTENDED. THERE WAS NO RESISTANCE FELT IN THE HANDLE WHEN ACTUATING THE CUPS. HOWEVER, THERE IS A GAP IN BETWEEN THE CUPS AND THE TEETH DO NOT MESH CORRECTLY. NO OTHER ANOMALIES WERE DETECTED WITH THE DEVICE. THE DEVICE WAS SENT BACK TO THE SUPPLIER FOR A FULL EVALUATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. THE SUPPLIER PROVIDED THE FOLLOWING: VISUAL EVALUATION: THE TIP BENT WAS NOT CONFIRMED. THE CUPS ARE MISALIGNED. FUNCTIONAL EVALUATION: WHEN THE HANDLE WAS ACTUATED THE TIP DID OPEN AND CLOSE. HOWEVER, THE CUPS WERE MISALIGNED SO NO FURTHER EVALUATION WAS PERFORMED. THE DEVICE HISTORY RECORDS FOR PACKAGING WORK ORDER (PWO) WERE REVIEWED. THE PWO CONSISTED OF ONE ASSEMBLY ORDER (AO). THE MANUFACTURING RECORDS AND/OR FINAL QUALITY CONTROL CHECKLIST DID INCLUDE RELEVANT DEFECTS. THERE WERE REJECTS FOR MISALIGNED CUPS IN THE ASSEMBLY ORDER. INVESTIGATION CONCLUSION: THE SUPPLIER PROVIDED THE FOLLOWING: THE ASSIGNABLE CAUSE OF THE MISALIGNED CUPS WAS NOT DETERMINED. THEREFORE, NO CORRECTIVE ACTION HAS BEEN ASSIGNED. PRIOR TO DISTRIBUTION, ALL COOK CAPTURA SERRATED FORCEPS W/O SPIKE ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE PROPER WORKABILITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED REMOTE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

IN PREPARATION FOR AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN SELECTED COOK CAPTURA SERRATED FORCEPS W/O SPIKE. THE PHYSICIAN OPENED THE PACKAGE AND DETECTED THE TIP OF THE DEVICE BENT AND THE CUPS WERE MISPLACED [MISALIGNED]. THERE WAS NO USE ON THE PATIENT. THE PHYSICIAN CHANGED TO ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.

Additional Manufacturer Narrative · 1

INITIAL REPORTER OCCUPATION: UNKNOWN. THE PRODUCT WAS RETURNED FOR EVALUATION AND THE INVESTIGATION IS ON-GOING. A FOLLOW-UP EMDR WILL BE PROVIDED WITHIN 30 DAYS OF SUBMISSION OF THIS REPORT WITH PRODUCT EVALUATION INFORMATION.

Description of Event or Problem · 1

IN PREPARATION FOR AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN SELECTED COOK CAPTURA SERRATED FORCEPS W/O SPIKE. THE PHYSICIAN OPENED THE PACKAGE AND DETECTED THE TIP OF THE DEVICE BENT AND THE CUPS WERE MISPLACED [MISALIGNED]. THERE WAS NO USE ON THE PATIENT. THE PHYSICIAN CHANGED TO ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.