FDA Adverse Event Injury Summary report: N

GREENFIELD

MDR report key: 9933147 · Received April 7, 2020

Report

Report Number
2134265-2020-04590
Event Type
Injury
Date Received
April 7, 2020
Date of Event
February 1, 2019
Report Date
July 28, 2020
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
DTK
PMA / PMN Number
K964284
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

ON (B)(6) 2004, THE PATIENT THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED A MESENTERIC PERFORATION, DEVICE TILT, AND STRUT BENDING. THERE WERE 13 DEGREES OF LEFT LATERAL TILT AND 12 DEGREES OF POSTERIOR TILT. SEVERAL OF THE STRUTS APPEARED BENT WITHOUT FRACTURE. ONE STRUT PERFORATED 3MM BEYOND THE INFERIOR VENA CAVA (IVC) WALL AND ANOTHER STRUT PERFORATED 2MM BEYOND THE IVC WALL. ON (B)(6) 2020, THE FILTER WAS REMOVED AS IT WAS MIGRATING TOWARDS THE KIDNEY. THE FILTER WAS INTACT UPON REMOVAL.

Additional Manufacturer Narrative · 1

PARTIAL LOT NUMBER OF 6864237 WAS REPORTED.

Description of Event or Problem · 1

ON (B)(6) 2004, THE PATIENT THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED A MESENTERIC PERFORATION, DEVICE TILT, AND STRUT BENDING. THERE WERE 13 DEGREES OF LEFT LATERAL TILT AND 12 DEGREES OF POSTERIOR TILT. SEVERAL OF THE STRUTS APPEARED BENT WITHOUT FRACTURE. ONE STRUT PERFORATED 3MM BEYOND THE INFERIOR VENA CAVA (IVC) WALL AND ANOTHER STRUT PERFORATED 2MM BEYOND THE IVC WALL. ON (B)(6) 2020, THE FILTER WAS REMOVED AS IT WAS MIGRATING TOWARDS THE KIDNEY. THE FILTER WAS INTACT UPON REMOVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396354 GREENFIELD FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK BOSTON SCIENTIFIC CORPORATION 43335 0006864237

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention