FDA Adverse Event Malfunction Summary report: N

ACCELERATOR A3600

MDR report key: 9932999 · Received April 7, 2020

Report

Report Number
3010825766-2020-00004
Event Type
Malfunction
Date Received
April 7, 2020
Report Date
April 7, 2020
Manufacturer
INPECO SA
Product Code
CEM
PMA / PMN Number
K121012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE EVENT HAS BEEN INVESTIGATED BY THE DISTRIBUTOR FSE, BUT IT HAS NOT BEEN POSSIBLE TO ESTABLISH THE CAUSE WHICH TRIGGERED THE FLAME. THE POWER SUPPLY IS CERTIFIED FOR THE ELECTRICAL SAFETY AND THE FLAME AUTO-EXTINGUISHED, AS EXPECTED. THE DAMAGED POWER SUPPLY HAS BEEN REPLACED. THERE ARE NO PREVIOUS SIMILAR OCCURRENCES REPORTED FROM THE FIELD FOR THIS COMPONENT, SO THE EVENT HAS BEEN EVALUATED AS A SINGLE OCCURRENCE. THE AUTOMATION SYSTEM IS NOW PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER ACTIONS ARE NEEDED.

Description of Event or Problem · 1

THE DISTRIBUTOR FSE ONSITE SMELLED SMOKE NEAR THE STORAGE RETRIEVE MODULE AND RIGHT AFTER THE TRACK WENT DOWN WITH A POWER SUPPLY ERROR. INVESTIGATING THE ISSUE, THE FSE FOUND THE 24V POWER SUPPLY DAMAGED WITH EVIDENCE OF FIRE. NO VISIBLE SMOKE NOR FLAMES WERE SEEN: THE FIRE WAS ALREADY AUTO EXTINGUISHED WHEN THE FSE FOUND THE DAMAGED POWER SUPPLY. NO ONE IN THE LABORATORY WAS NEAR TO THE POWER SUPPLY WHEN THE EVENT OCCURRED AND NO ONE GOT INJURED. NO OTHER COMPONENTS WERE DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396772 ACCELERATOR A3600 LABORATORY AUTOMATION SYSTEM CEM INPECO SA

Patients

Seq Age Sex Outcome Treatment
1