FDA Adverse Event Injury Summary report: N

ACCUFIX

MDR report key: 99205 · Received June 17, 1997

Report

Report Number
1723248-1997-00474
Event Type
Injury
Date Received
June 17, 1997
Date of Event
May 13, 1997
Report Date
June 16, 1997
Manufacturer
TELECTRONICS PACING SYSTEMS
Product Code
DTB
Removal / Correction Number
Z-2095
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION UNDER 30X MAGNIFICATION INDICATES NO J WIRE FRACTURES. X-RAY OF LEAD CONFIRMS NO J STIFFENER WIRE FRACTURES.

Description of Event or Problem · 1

DEVICE ANALYSIS IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCUFIX Implant PERM IMPL PACER ELECTRODE DTB TELECTRONICS PACING SYSTEMS 330-801 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention