TISSUE-TEK CRYO3 FLEX, 115V D
Report
- Report Number
- 2083544-2020-00002
- Event Type
- Malfunction
- Date Received
- April 1, 2020
- Date of Event
- February 10, 2020
- Report Date
- July 8, 2020
- Manufacturer
- SAKURA FINETEK USA, INC.
- Product Code
- IDO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE INVESTIGATION OF THE RETURNED UNIT REVEALED THAT THE SSR (SOLID STATE RELAY) WAS SLOWLY OVERHEATED AND EVENTUALLY BURNED. THE SURROUNDING COMPONENTS, SUCH AS AIR COMPRESSOR, COOLING FAN, CIRCUIT BREAKER, AC LINE FILTER, ETC., WHICH MAY HAVE DIRECT OR INDIRECT EFFECT ON THE SSR'S PERFORMANCE WERE EVALUATED; NO ISSUE WAS IDENTIFIED, THUS, ELIMINATED AS A CONTRIBUTING FACTOR OR CAUSE OF THE EVENT. THE INVESTIGATION OF THE LOG FILE REVEALED THAT THE SSR (SOLID STATE RELAY), WHICH NORMALLY TURNS ON/OFF THE AC INPUT TO THE COMPRESSOR, HAD AN INTERNAL ISSUE AS INDICATED ON THE LOG FILE SINCE 1/19/20. IT WAS NOT ABLE TO TURN OFF THE COMPRESSOR WHEN THE COMPRESSOR WAS COMMANDED TO DO SO BY THE SW. THE SSR WAS OVERHEATED AND EVENTUALLY BURNED, THUS, SAKURA WAS NOT ABLE TO DETERMINE EXACTLY WHAT INTERNAL ISSUE THE SSR HAD. THE DEVICE WAS DESIGNED, TESTED AND CERTIFIED BY INTERTEK, A GLOBAL PRODUCT QUALITY TESTING ORGANIZATION, TO BE IN COMPLIANCE WITH THE IEC 61010-1:2010, WHICH INCLUDES THE REQUIREMENTS IN CLAUSE 9 FOR PROTECTION AGAINST THE SPREAD OF FIRE. THE DAMAGE WAS WHOLLY CONTAINED WITHIN THE INSIDE OF THE DEVICE AS DESIGNED. IF THIS MALFUNCTION OCCURS, THE FIRE IS CONTAINED WITHIN THE MAIN POWER BOX AND WILL NOT IMPACT OTHER PARTS OF THE DEVICE. THIS IS CONSIDERED AN ISOLATED COMPONENT FAILURE. THIS IS THE ONLY INCIDENT OF THIS TYPE SINCE THE LAUNCH OF THIS DEVICE IN 2016. (B)(4).
SAKURA FINETEK USA, INC., RECEIVED A COMPLAINT ON (B)(6) 2020 IN WHICH CUSTOMER STATED THAT A SMELL OF SMOKE CAME OUT FROM THE BACK OF THE UNIT ON (B)(6) 2020 WHEN IT WAS IDLE AND NOT IN USE. THE TECHNICIAN IN THE LAB SHUT THE UNIT OFF AND AIRED OUT THE ROOM. THIS CAUSED THE SSR (SOLID STATE RELAY) AND RIBBON CABLES TO SEVERELY BURN AND MELT. THE INCIDENT OCCURRED AFTER THE DAILY AUTO OZONE DISINFECTION CYCLE ENDED WHILE THE DEVICE WAS NOT IN USE BY THE OPERATOR, AND NO PATIENT TISSUES WERE DAMAGED AS A RESULT OF THIS INCIDENT.
THE SAKURA INVESTIGATION OF THE LOG FILE REVEALED THAT THE SSR (SOLID STATE RELAY), WHICH NORMALLY TURNS ON/OFF THE AC INPUT TO THE COMPRESSOR, HAD AN INTERNAL ISSUE AS INDICATED ON THE LOG FILE SINCE 1/19/2020. IT WAS NOT ABLE TO TURN OFF THE COMPRESSOR WHEN IT WAS COMMANDED TO DO SO BY THE SW. THE SSR WAS OVERHEATED AND EVENTUALLY BURNED, THUS, SAKURA IS NOT ABLE TO DETERMINE EXACTLY WHAT INTERNAL ISSUE THE SSR HAD. THE INCIDENT OCCURRED AFTER THE DAILY AUTO OZONE DISINFECTION CYCLE ENDED WHILE THE DEVICE WAS NOT IN USE BY THE OPERATOR, AND NO PATIENT TISSUES WERE DAMAGED AS A RESULT OF THIS INCIDENT. THE DEVICE WAS DESIGNED TO MEET THE REQUIREMENTS OF THE IEC 61010-1:2010, SPECIFICALLY THE REQUIREMENT OF CLAUSE 9 PROTECTION AGAINST THE SPREAD OF FIRE. THE FIRE/FLAME DID NOT SPREAD OUTSIDE OF THE DEVICE, WAS CONTAINED WITHIN THE INSIDE OF THE DEVICE AS DESIGNED. THIS IS CONSIDERED AN ISOLATED COMPONENT FAILURE; SAKURA HAS NOT SEEN SUCH FAILURE SINCE THE CRYO3 FLEX WAS INTRODUCED IN 2016.
SAKURA FINETEK USA, INC., RECEIVED A COMPLAINT ON 02/11/2020 IN WHICH CUSTOMER STATED THAT UNIT STARTED ON FIRE FROM THE BACK ON (B)(6) 2020 WHEN IT WAS IDLE AND NOT IN USE. THE TECHNICIAN IN THE LAB SHUT THE UNIT OFF AND AIRED OUT THE ROOM. THE FIRE CAUSED THE SSR (SOLID STATE RELAY) AND RIBBON CABLES TO SEVERELY BURN AND MELT. THERE WAS NO DAMAGE TO THE PATIENT TISSUE SAMPLES, OR INJURY TO THE USER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372562 | TISSUE-TEK CRYO3 FLEX, 115V D | CRYO3 FLEX | IDO | SAKURA FINETEK USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |