FDA Adverse Event Injury Summary report: N

HEALON GV PRO

MDR report key: 9901381 · Received March 30, 2020

Report

Report Number
3004750704-2020-00025
Event Type
Injury
Date Received
March 30, 2020
Date of Event
December 5, 2019
Report Date
October 30, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
LZP
UDI-DI
05050474651593
PMA / PMN Number
P810031
Removal / Correction Number
2020664-12/02/19-001-R
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE (B)(4) AND (B)(4) .

Additional Manufacturer Narrative · 1

AGE/DATE OF BIRTH: UNKNOWN, INFORMATION NOT PROVIDED. GENDER/SEX: UNKNOWN, INFORMATION NOT PROVIDED. IF IMPLANTED; GIVE DATE: N/A (NOT APPLICABLE). HEALON IS NOT AN IMPLANTABLE DEVICE. IF EXPLANTED; GIVE DATE: N/A (NOT APPLICABLE). HEALON IS NOT AN IMPLANTABLE DEVICE; THEREFORE, NOT EXPLANTED. LAST NAME: UNKNOWN, INFORMATION NOT PROVIDED. (B)(6). (B)(4). THIS COMPLAINT IS PART OF THE RECALL - REPORT NUMBER 2020664-12/02/19-001-R: JOHNSON & JOHNSON SURGICAL VISION (JJSV) ISSUED A VOLUNTARY RECALL ON NOVEMBER 22, 2019. IT HAS BEEN REPORTED CUSTOMERS HAVE DESCRIBED HEALON GV PRO AS BEHAVING DIFFERENTLY THAN THE LEGACY HEALON GV, ESPECIALLY IN REGARD TO THE TECHNIQUES REQUIRED TO REMOVE THE PRODUCT FROM THE EYE. AN INCREASE OF INTRA-OCULAR PRESSURE (IOP) IS REPORTED IF THERE ARE SMALL AMOUNTS OF HEALON GV PRO REMAINING BEHIND THE OPERATIVE EYE. THIS VOLUNTARY RECALL IS BEING INITIATED DUE TO RECEIVED REPORTS OF HEALON GV PRO BEING DIFFICULT TO REMOVE FROM THE EYE, LEADING TO INCREASED POST-OPERATIVE IOP REQUIRING ADDITIONAL INTERVENTION. POTENTIAL CLOGGING OF PHACOEMULSIFICATION EQUIPMENT TUBING HAS ALSO BEEN REPORTED, WHICH MAY LEAD TO DELAY IN THE PROCEDURE OR OCULAR INJURY. THERE ARE TWENTY-ONE AFFECTED LOT NUMBERS. THE RECALL NOTIFICATION LETTER HAS BEEN SENT TO ALL CUSTOMERS INSTRUCTING THEM TO RETURN THE UNITS OF HEALON GV PRO FROM THE TWENTY-ONE (21) AFFECTED LOTS. JOHNSON & JOHNSON SURGICAL VISION HAS INITIATED A CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) TO INVESTIGATE AND ADDRESS THE ISSUE. ACTION ITEMS GENERATED FROM THE CAPA WILL BE SUBMITTED IN THE FUTURE INTERIM REPORT(S) AS PART OF THE RECALL PROCESS. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE INJECTION OF THE HEALON GV PRO, THE SURGEON NOTED THE PRODUCT WAS DIFFICULT TO EXPEL AND THAT IT CLOGGED THE HAND PIECE. THE SURGEON REPORTED OF ACUTE AND CHRONIC POST OPERATION HYPERTONIA LEADING TO ABNORMALITIES, POSTOP REFRACTION CORNEAL BURN, DEFORMED PUPIL, INTRAOCULAR PRESSURE INCREASES, LENS DISLOCATION, SWELLING OF THE CAPSULAR BAG LEADING TO POST REFRACTION ANOMALIES, BLINDNESS IN A CASE OF GLAUCOMA PATIENT AND HYPERTONIA REQUIRING AN EARLY YTTRIUM-ALUMINUM GARNET (YAG) TREATMENTS. REPORTEDLY, ABOUT ONE HUNDRED (100) PATIENTS ARE AFFECTED. SYMPTOMS WERE TREATED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT CAPTURES THE EVENT FOR FEW PATIENTS HAVING SOME PAIN AND REQUESTED ADDITIONAL CONSULTATION. HYPERTONIA WAS REPORTED, MEASURED ROUTINELY BUT APPROXIMATELY 35-45MMHG THEN LOWERED TO 25-28MMHG POST INTERVENTION. THE BAG WAS WASHED OF BUT NOT ENOUGH. BAD REFRACTION ERRORS. FOR 1 CASE THE LENS WAS FOUND DISLOCATED, DUE TO ON AMOUNT OF OVD LEFT IN THE BAG. THE LENS WAS PUSHED FORWARD AND REQUIRED AN ADDITIONAL INTERVENTION TO REMOVE THE OVD. YAG WAS DONE A MONTH AFTER INTERVENTION. NONE OF THE PATIENTS EXPERIENCED INCREASED INTRAOCULAR PRESSURE POST-OPERATIVELY. ADDITIONALLY, THE CUSTOMER PROVIDED TWO DIFFERENT LOT NUMBERS. DUE THE LIMITED INFORMATION AVAILABLE TO DATE, IT WAS NOT POSSIBLE TO OBTAIN THE BATCH NUMBER RELATED TO THIS SPECIFIC CASE. THIS REPORT WILL PERTAINS TO THE HEALON GV PRO, LOT NUMBER UE31283. A SEPARATE REPORT WILL BE SUBMITTED ON THE LOT NUMBER UE31098.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
363542 HEALON GV PRO OVDS LZP JOHNSON & JOHNSON SURGICAL VISION, INC. TG85ML UE31283 05050474651593

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention