FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 9897221 · Received March 30, 2020

Report

Report Number
3004464228-2020-04575
Event Type
Malfunction
Date Received
March 30, 2020
Date of Event
March 17, 2020
Report Date
March 19, 2020
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K162296
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED AND EVALUATED. THE DOWNLOAD DATA REVEALED THAT THE POD WAS IN POD STATE 14, INDICATING THAT POD ACTIVATION WAS NOT COMPLETED AFTER IT WAS FILLED. THE POD COULD NOT BE ACTIVATED IN THIS STATE AND THUS, BOTH THE NEEDLE DEPLOYMENT AND FLUID DELIVERY WERE PROHIBITED.THE PRODUCT WAS RETURNED WITH A DIFFERENT LOT NUMBER THAN WAS REPORTED AND NOTED ON THE INITIAL MDR. CORRECTION TO D(4): LOT NUMBER CHANGED FROM UNAVAILABLE TO L45179. EXPIRATION DATE CHANGED FROM UNAVAILABLE TO 4/8/2021. MODEL NO CHANGED FROM 14810 TO 19191. CORRECTION TO G(5): PMA/510(K) # CHANGED FROM K122953 TO K162296. UNIQUE IDENTIFIER (UDI) # CHANGED FROM UNAVAILABLE TO (B)(4). CORRECTION TO H(4): DEVICE MFG DATE CHANGED FROM UNAVAILABLE TO 10/8/2019.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. NO LOT RELEASE RECORDS WERE REVIEWED, AS THE PRODUCT LOT NUMBER WAS NOT PROVIDED. MYLIFE OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE. MODEL: ENT450. 14518-5C-AW REV E 03/16. USING THE POD 5 / PAGE 53. WARNING: CHECK THE INFUSION SITE AFTER INSERTION TO ENSURE THAT THE CANNULA WAS PROPERLY INSERTED. YOU SHOULD CHECK YOUR BLOOD GLUCOSE 1.5 TO 2 HOURS AFTER EACH POD CHANGE AND CHECK THE INFUSION SITE PERIODICALLY. IF THE CANNULA IS NOT PROPERLY INSERTED, HYPERGLYCEMIA MAY RESULT. VERIFY THERE IS NO WETNESS OR SCENT OF INSULIN, WHERE AS MAY INDICATE THE CANNULA HAS DISLODGED.

Description of Event or Problem · 1

IT WAS REPORTED A NEEDLE MECHANISM FAILURE OCCURRED. THE POD WAS NOT WORN AND NO ADVERSE PATIENT IMPACT WAS REPORTED.