FDA Adverse Event Injury Summary report: N

STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 3

MDR report key: 9896985 · Received March 30, 2020

Report

Report Number
3005180920-2020-00203
Event Type
Injury
Date Received
March 30, 2020
Date of Event
March 4, 2020
Report Date
March 30, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802041
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 13 MARCH 2020: LOT 163265: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02-SEP-2016. EXPIRATION DATE: 2021-08-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR EVENT REPORTED. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: PARTIAL HIP REVISION SURGERY (STEM AND HEAD) PERFORMED 3 YEARS AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD FARMER. RADIOLOGICAL IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES IN GRUEN ZONES 1, 2, 6 AND 7 AND SIGNS OF STRESS SHIELDING SUGGESTING STEM MOBILIZATION. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY AND CAUSES ARE OFTEN UNKNOWN. EARLY RETURN TO HIGH LEVEL OF ACTIVITIES MAY REDUCE PRIMARY STABILITY OF THE STEM AND HINDER IMPLANT FIXATION. NO CONCLUSION CAN BE DRAWN ON THE BASIS OF AVAILABLE INFORMATION.

Description of Event or Problem · 1

REVISION SURGERY PERFORMED 3 YEARS AND 1 MONTH AFTER THE PRIMARY SURGERY DUE TO THE PROXIMAL LOOSENING OF A QUADRA H STEM. THE SURGEON REVISED THE STEM AND BALL HEAD. THE PATIENT WAS A HEAVY WORKING FARMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359797 STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 3 CEMENTLESS HIP STEM JDI MEDACTA INTERNATIONAL SA 01.12.023 163265 07630030802041

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention