MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2020-04939
- Event Type
- Injury
- Date Received
- March 27, 2020
- Date of Event
- January 27, 2020
- Report Date
- March 18, 2020
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000297
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
ON MAY 6, 2020, THE PRODUCT INVESTIGATION WAS COMPLETED. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED FOR THE COMPLAINT DEVICE. THE MANUFACTURING RECORD EVALUATION (MRE) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. DEVICE INVESTIGATION SUMMARY: DURING VISUAL EVALUATION, AN ANOMALY WAS OBSERVED ON THE DEVICE CONSISTENT WITH A CREASE/FOLD IN THE POSTERIOR VIEW . A TEAR WAS ALSO OBSERVED WITHIN THE CREASE/FOLD, MEASURING 2.0 CM. THE EVALUATION DETERMINED THAT THE LOSS OF SHELL INTEGRITY IS CONSISTENT WITH A CREASE FOLD RUPTURE OF THE IMPLANT SHELL, WHICH IS A KNOWN INHERENT RISK OF GEL-FILLED MAMMARY PROSTHESIS. RUPTURE CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN A RUPTURE IN A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF THIS 5983295 LOT-SERIAL NUMBER WERE REVIEWED AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS LOT. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2020, MENTOR BECAME AWARE THAT THE DEVICE THAT WAS INITIALLY RETURNED FOR THE CONTRALATERAL SIDE (RIGHT SIDE) REPORTED UNDER MRN 1645337-2020-04936 INITIAL REPORT, WAS ACTUALLY THE LEFT SIDE IMPLANT. THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4). THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE, MATERIAL RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) YEAR OLD FEMALE PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 325CC MENTOR MEMORYGEL BREAST IMPLANTS, SUFFERED BILATERAL BREAST CAPSULAR CONTRACTURE; BAKER GRADE IV, POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT BILATERAL IMPLANT EXPLANTATION WITHOUT REPLACEMENT ON (B)(6) 2020. DURING THE EXPLANTATION SURGERY, IT WAS DISCOVERED THAT THE LEFT BREAST IMPLANT WAS RUPTURED. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 354809 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 3503251BC | 5983295 | 00081317000297 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR | Required Intervention |