FDA Adverse Event Malfunction Summary report: N

THIN OSTEOTOME HANDLE

MDR report key: 9883441 · Received March 26, 2020

Report

Report Number
1818910-2020-09302
Event Type
Malfunction
Date Received
March 26, 2020
Date of Event
March 10, 2020
Report Date
March 10, 2020
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
HWM
UDI-DI
10603295123071
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: THE DEVICE WAS REVIEWED BY DEPUY ENGINEERING MELBOURNE IN A-3910355 WHICH STATES MY ASSESSMENT IS THAT THIS COMPLAINT IS DUE TO WEAR AND TEAR. ROOT CAUSE ATTRIBUTED TO THE DEVICE BEING WORN FROM NORMAL USE AND SERVICING. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: THE DEVICE WAS REVIEWED BY DEPUY ENGINEERING MELBOURNE IN A-3910355 WHICH STATES MY ASSESSMENT IS THAT THIS COMPLAINT IS DUE TO WEAR AND TEAR. ROOT CAUSE ATTRIBUTED TO THE DEVICE BEING WORN FROM NORMAL USE AND SERVICING. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT #: (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THEATRE STAFF REPORTED OSTEOTOME HANDLE IN CORAIL SET IS NOT HOLDING THE OSTEOTOME BLADES, PLEASE REPLACE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347918 THIN OSTEOTOME HANDLE HIP INSTRUMENTS : HANDLES HWM DEPUY ORTHOPAEDICS INC US 2422-10-000 VZ14948 10603295123071

Patients

Seq Age Sex Outcome Treatment
1