ACCU-CHEK ® INFORM II TEST STRIPS
Report
- Report Number
- 1823260-2020-00846
- Event Type
- Malfunction
- Date Received
- March 26, 2020
- Date of Event
- March 2, 2020
- Report Date
- March 26, 2020
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- LFR
- PMA / PMN Number
- K121679
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IT WAS NOTED THAT STRIP LOT 477954 WAS USED ON THE FLOOR BY THE STAFF AND STRIP LOT 477716 WAS USED BY THE CUSTOMER N HER OFFICE TO PERFORM CONTROL TESTS. IT WAS VERIFIED THAT BOTH LEVEL OF CONTROLS WERE RAN ON METER (B)(4) PRIOR TO THE QUESTIONABLE RESULTS AND PASSED. A HIGH LEVEL CONTROL WAS RAN AFTER THE QUESTIONABLE RESULTS WHICH FAILED TWICE ON METER (B)(4). USING METER (B)(4) THE CUSTOMER RAN BOTH LOW LEVEL AND HIGH LEVEL CONTROLS AND BOTH FAILED. LOW LEVEL CONTROL RANGE IS 30MG/DL - 60MG/DL. HIGH LEVEL CONTROL RANGE IS 261MG/DL - 353MG/DL. THE CUSTOMER STATED THEY GOT A RESULT OF LO AND 19MG/DL WITH LOW LEVEL CONTROL AND 268MG/DL WITH HIGH LEVEL CONTROL. THE CUSTOMER'S METERS, TEST STRIPS (LOT 477716, 4 STRIPS) AND CONTROLS WERE PROVIDED FOR INVESTIGATION. STRIP LOT 477954 WAS NOT RETURNED FOR INVESTIGATION. TESTING RESULTS: (B)(4). LEVEL 1 QC 1: 46 MG/DL, LEVEL 1 QC 2: 45 MG/DL, LEVEL 1 QC 3: 46 MG/DL, LEVEL 2 QC 1: 310 MG/DL, LEVEL 2 QC 2: 317 MG/DL, LEVEL 2 QC 3: 312 MG/DL. (B)(4). LEVEL 1 QC 1: 45 MG/DL, LEVEL 1 QC 2: 45 MG/DL, LEVEL 1 QC 3: 45 MG/DL, LEVEL 2 QC 1: 306 MG/DL, LEVEL 2 QC 2: 309 MG/DL, LEVEL 2 QC 3: 309 MG/DL. THE OBTAINED QC VALUES WERE IN THE ALLOWED RANGE OF THE USED COMBINATION STRIP LOT - QC LOT. ALL MEASUREMENTS WERE WITHOUT ERROR MESSAGES. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
THE INITIAL REPORTER COMPLAINED OF A DISCREPANT GLUCOSE RESULT WITH ACCU-CHEK INFORM II METER SERIAL NUMBER (B)(4) WHEN COMPARED TO AN ACCU-CHEK INFORM II METER SERIAL NUMBER (B)(4). AT 8:53AM A HEELSTICK WAS PERFORMED ON INFORM II (B)(4). THE RESULT WAS 23MG/DL. AT 8:57AM A HEELSTICK WAS PERFORMED ON INFORM II (B)(4). THE RESULT WAS 48MG/DL. THE CUSTOMER DID NOT BELIEVE THE RESULTS OBTAINED ON METER (B)(4). THE PATIENT IS CURRENTLY FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 346901 | ACCU-CHEK ® INFORM II TEST STRIPS | BLOOD GLUCOSE MONITORING TEST STRIPS | LFR | ROCHE DIAGNOSTICS | NA | 477954 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 MO |