FDA Adverse Event Death Summary report: N

LUCAS¿ CPR CHEST COMPRESSION SYSTEM

MDR report key: 9881681 · Received March 25, 2020

Report

Report Number
3005445717-2020-00003
Event Type
Death
Date Received
March 25, 2020
Date of Event
August 5, 2019
Report Date
March 30, 2020
Manufacturer
JOLIFE AB - 3005445717
Product Code
DRM
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
GB
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

F6 USER FACILITY AWARENESS DATE OF THE INITIAL SUBMISSION WAS LEFT BLANK. F6 USER FACILITY AWARENESS DATE SHOULD BE 02/25/2020.

Description of Event or Problem · 0

THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT WORK DURING A PATIENT EVENT. THE DEVICE INDICATED THE BATTERY WAS FULLY CHARGED, HOWEVER THE DEVICE DID NOT OPERATE. THE DEVICE'S SUCTION CUP WAS PLACED ON THE PATIENT, BUT THE DEVICE WOULD NOT OPERATE WHEN THE CONTINUOUS AND 30:2 BUTTONS WERE PRESSED. EVENTUALLY, THE DEVICE WAS REMOVED FROM THE PATIENT AND MANUAL CPR WAS PROVIDED. WHEN THE DEVICE WOULD NOT OPERATE AND IT WAS REMOVED FROM THE PATIENT, IT WAS PLUGGED INTO THE POWER SUPPLY, TO CHARGE THE BATTERY FOR 10 MINUTES. THE DEVICE SUBSEQUENTLY OPERATED CORRECTLY. THE INTERRUPTION IN CPR WAS ESTIMATED TO BE APPROXIMATELY 1 MINUTE. IT WAS DELAYED DUE TO DECISION MAKING AROUND TRANSPORTING WITHOUT A LUCAS, WHILE IT WAS BEING CHARGED. THE PATIENT DID NOT SURVIVE THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL CONTACTED THE CUSTOMER TO REQUEST ADDITIONAL INFORMATION ON THE PATIENT. THE CUSTOMER PROVIDED PHYSIO-CONTROL WITH ALL THE AVAILABLE PATIENT INFORMATION. PATIENT FIELDS IN WHICH INFORMATION IS NOT PROVIDED WERE INTENTIONALLY LEFT BLANK. STRYKER COMPLETED A CLINICAL REVIEW AND DETERMINED THAT THE DEVICE USE MAY HAVE CONTRIBUTED TO THE PATIENT OUTCOME, AS THE INTERRUPTION IN CPR WAS ONE MINUTE. THE USERS DID NOT FOLLOW THE INSTRUCTIONS FOR USE THAT STATE IF THERE ARE INTERRUPTIONS, TO REMOVE THE DEVICE AND START MANUAL CHEST COMPRESSIONS. BY FOLLOWING THE INSTRUCTIONS FOR USE THE INTERRUPTION IN CPR IS LIKELY TO BE SHORT AND NOT, AS IN THIS CASE, AS LONG AS ONE MINUTE. STRYKER EVALUATED THE CUSTOMER'S DEVICE AND WAS UNABLE TO DUPLICATE THE REPORTED ISSUE. AFTER PERFORMING AN UNRELATED REPAIR, PROPER DEVICE OPERATION WAS OBSERVED. THE DEVICE WAS SUBSEQUENTLY RETURNED TO THE CUSTOMER FOR USE. IT WAS NOTED THAT THE BATTERIES USED WITH THE DEVICE WERE MANUFACTURED IN 2015. STRYKER RECOMMENDS THAT BATTERIES BE REPLACED APPROXIMATELY EVERY THREE YEARS. THE LIKELY CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE DUE TO USER ERROR, AS THE CUSTOMER HAD NOT REPLACED THE BATTERY AFTER THE RECOMMENDED 3 YEAR LIFE. INITIAL MDR MFR 0003015876-2019-01465 WAS ORIGINALLY SUBMITTED ON 09/03/2019 BASED ON AWARENESS DATE (B)(6) 2019. THE DEVICE EVALUATION WAS COMPLETED ON (B)(6) 2020 PROMPTING A CORRECTION AND THIS ADDITIONAL REPORT.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT WORK DURING A PATIENT EVENT. THE DEVICE INDICATED THE BATTERY WAS FULLY CHARGED, HOWEVER THE DEVICE DID NOT OPERATE. THE DEVICE'S SUCTION CUP WAS PLACED ON THE PATIENT, BUT THE DEVICE WOULD NOT OPERATE WHEN THE CONTINUOUS AND 30:2 BUTTONS WERE PRESSED. EVENTUALLY, THE DEVICE WAS REMOVED FROM THE PATIENT AND MANUAL CPR WAS PROVIDED. WHEN THE DEVICE WOULD NOT OPERATE AND IT WAS REMOVED FROM THE PATIENT, IT WAS PLUGGED INTO THE POWER SUPPLY, TO CHARGE THE BATTERY FOR 10 MINUTES. THE DEVICE SUBSEQUENTLY OPERATED CORRECTLY. THE INTERRUPTION IN CPR WAS ESTIMATED TO BE APPROXIMATELY 1 MINUTE. IT WAS DELAYED DUE TO DECISION MAKING AROUND TRANSPORTING WITHOUT A LUCAS, WHILE IT WAS BEING CHARGED. THE PATIENT DID NOT SURVIVE THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342453 LUCAS¿ CPR CHEST COMPRESSION SYSTEM COMPRESSOR, CARDIAC, EXTERNAL DRM JOLIFE AB - 3005445717 LUCAS 3015C620

Patients

Seq Age Sex Outcome Treatment
1 71 YR Death