FDA Adverse Event Malfunction Summary report: N

G7 STR INSRTR THREADED SHAFT

MDR report key: 9879996 · Received March 25, 2020

Report

Report Number
0001825034-2020-01274
Event Type
Malfunction
Date Received
March 25, 2020
Date of Event
February 18, 2020
Report Date
July 30, 2020
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
K121874
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

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

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 110010248, G7 OSSEOTI 4 HOLE SHELL 60MM G, LOT NUMBER 6665675, 010002736, G7 BALL HEX DRVR FOR INSR HNDL, LOT NUMBER 061706, 110003451, G7 STR MODULAR SHELL INSERTER, LOT NUMBER 253370. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2020-01275, 0001825034-2020-01277.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SCRUB TECH UNKNOWINGLY CROSS THREADED THE G7 STRAIGHT INSERTER ONTO THE G7 CUP DURING INITIAL THA. UPON IMPACTING THE CUP INTO THE ACETABULUM, THE INSERTER BECAME COLD WELDED TO THE G7 CUP. INSERTER SCREWDRIVER BALL TIP BROKE OFF WHILE TRYING TO DISENGAGE INSERTER FROM CUP. CASE WAS CONVERTED TO THE CONTINUUM ACETABULAR SYSTEM WITH NO IMPACT TO THE SURGEON OR THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343641 G7 STR INSRTR THREADED SHAFT ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LPH ZIMMER BIOMET, INC. N/A 061289

Patients

Seq Age Sex Outcome Treatment
1