INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2020-04496
- Event Type
- Injury
- Date Received
- March 25, 2020
- Date of Event
- February 18, 2020
- Report Date
- July 8, 2020
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628007352
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: DEVICE PHOTOGRAPH(S) FOR THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS REVIEWED ON 26-JUN-20. VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED: DEVICE PATCH WITH LOT NUMBER 3222503, CLOUDY IN THE GEL, IT IS LABELED AS RIGHT SIDE AND APPARENTLY HAS NO OPENINGS. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. REFER TO PS-QP-ONEV-115717: COVID-19 QUALITY PLAN - COMPLAINT HANDLING. ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B.5, D.7, H.6.
DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF "CAPSULAR CONTRACTURE" IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE III.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343788 | INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | SRF-485 | 3222503 | 10888628007352 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |