STEM: AMISTEM H HA COATED STD STEM SIZE 4
Report
- Report Number
- 3005180920-2020-00174
- Event Type
- Injury
- Date Received
- March 25, 2020
- Date of Event
- February 25, 2020
- Report Date
- March 25, 2020
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030804106
- PMA / PMN Number
- K093944
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 6 MARCH 2020: LOT 140758: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28-APR-2014. EXPIRATION DATE: 2019-03-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR EVENTS REPORTED. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR 5 YEARS AFTER PRIMARY CEMENTLESS THA THE STEM IS REVISED DUE TO MOBILIZATION. THE LOW-QUALITY IMAGES SHOW AN OSTEOLYTIC REGION IN ZONE 6 OF UNKNOWN ORIGIN. NO COMMENTS WERE MADE TO THE STATUS OF THE ACETABULAR INSERT. SOME SIGNS OF RADIOGRAPHIC LOOSENING ARE LASO VISIBLE IN THE PROXIMAL REGION OF THE FEMUR. THE MAIN CAUSE FOR THIS LOOSENING CANNOT BE DETERMINED WITH THE INFORMATION AT HAND.
LOOSENING OF AN AMISTEM-H 5 YEARS AFTER THE PRIMARY SURGERY. THE SURGEON REVISED THE STEM LINER AND HEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342141 | STEM: AMISTEM H HA COATED STD STEM SIZE 4 | CEMENTLESS HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 01.18.134 | 140758 | 07630030804106 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention |