FDA Adverse Event Malfunction Summary report: N

SYRINGE PLASTIPAK 20ML LL S/SU

MDR report key: 9872403 · Received March 24, 2020

Report

Report Number
3003916417-2020-00096
Event Type
Malfunction
Date Received
March 24, 2020
Date of Event
February 21, 2020
Report Date
April 17, 2020
Manufacturer
BECTON DICKINSON IND. CIRURGICAS LTDA
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D.10 DEVICE AVAILABLE FOR EVAL YES, RETURNED TO MANUFACTURER ON: 2020-03-30. H.6. INVESTIGATION SUMMARY : A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9326902, QUALITY NOTIFICATION AND MAINTENANCE ANALYSIS AND NO OCCURRENCES POTENTIALLY RELATED TO THE DEFECT WAS OBSERVED. THE SAMPLE SENT BY THE CUSTOMER WAS VERIFIED AND IT WAS POSSIBLE OBSERVE FOREIGN MATTER ¿ DIRT. THIS OCCURRENCE IS RELATED TO A CONTAMINATION DURING PRODUCTION PROCESS CAUSED BY PERSONNEL FAILURE DURING THE PACKAGING LINE CLEARANCE. THE OPERATORS WILL BE NOTIFIED FROM THIS COMPLAINT. THE DEFECT IDENTIFIED FROM THIS COMPLAINT WILL BE MONITORED FOR TREND EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE PLASTIPAK 20ML LL S/SU CONTAINED FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DIRTY IN THE SYRINGE LUER. ADDITIONAL INFORMATION: LOT NUMBER: 9326902; MATERIAL NUMBER: 990687. THE INCIDENT WAS NOTICED BEFORE THE USE. ANVISA NOTIFICATION: 2020.03.000455".

Additional Manufacturer Narrative · 1

INITIAL REPORTER ADDITIONAL PHONE #: (B)(6). INITIAL REPORTER FACILITY NAME: (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE PLASTIPAK 20ML LL S/SU CONTAINED FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DIRTY IN THE SYRINGE LUER. ADDITIONAL INFORMATION: LOT NUMBER: 9326902; MATERIAL NUMBER: (B)(4). THE INCIDENT WAS NOTICED BEFORE THE USE. ANVISA NOTIFICATION: (B)(4)".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339630 SYRINGE PLASTIPAK 20ML LL S/SU SYRINGE FMF BECTON DICKINSON IND. CIRURGICAS LTDA 9326902

Patients

Seq Age Sex Outcome Treatment
1 Other