3.5MM LOCKING CORTICAL SCREW WITH T15 HEXALOBE, SIZE 26MM
Report
- Report Number
- 3006460162-2020-00045
- Event Type
- Malfunction
- Date Received
- March 20, 2020
- Report Date
- May 28, 2020
- Manufacturer
- ORTHOPEDIATRICS, INC
- Product Code
- HWC
- UDI-DI
- 00841132110881
- PMA / PMN Number
- K100240
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4: DATE OF THIS REPORT: 5/28/2020. G7: TYPE OF REPORT: FOLLOW-UP. H3: DEVICE EVALUATED: YES. H2: FOLLOW-UP TYPE: ADDITIONAL INFORMATION, DEVICE EVALUATION. H4: DEVICE MANUFACTURE DATE: 20-APRIL-2017. H6: EVENT PROBLEM AND EVALUATION CODES: METHOD CODE(S): 3331 - 10 - 4110. RESULT CODE(S): 3252. CONCLUSION CODE(S): 22 - 4315. SCREW WAS EVALUATED AND THE REPORTED FRACTURE WAS CONFIRMED. A DHR REVIEW WAS CONDUCTED AND NO MANUFACTURING DESCREPANCIES WERE IDENTIFIED. A REVIEW OF THE APPLICABLE RISK MANAGMENT FILE WAS CONDUCTED AND IDENTIFIED BENDING AND OR FRACTURING POST-OPERATIVELY. THIS RISK FALLS WITHIN THE ACCEPTANCE CRITERIA. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ORTHOPEDIATRICS WILL CONTINUE TO MONITOR FOR TRENDS.
(B)(4). CONCOMITANT DEVICES: ITEM: 00-0903-2522, 3.5MM SELF TAPPING CORTICAL SCREW WITH T15 HEXALOBE, SIZE 22MM, LOT: 178767-J; ITEM: 00-0907-3760, 3.5MM CANNULATED LOCKING SCREW, 60MM, LOT: M72783-D; ITEM: 00-0907-3760, 3.5MM CANNULATED LOCKING SCREW, 60MM, LOT: M72783-D; ITEM: 00-0907-3755, 3.5MM CANNULATED LOCKING SCREW, 55MM, LOT: M70765-D; ITEM: 00-0903-2528, 3.5MM SELF TAPPING CORTICAL SCREW WITH T15 HEXALOBE, SIZE 28MM, LOT: 171316-J; ITEM: 00-0903-2524, 3.5MM SELF TAPPING CORTICAL SCREW WITH T15 HEXALOBE, SIZE 24MM, LOT: 178768-J; ITEM: 00-0907-2104, 130° X 3.5MM LPF PLATE X 4 HOLE, LOT: 046734-G. THE CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN THE PROCESS OF RETURNING TO ORTHOPEDIATRICS FOR EVALUATION. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 30064660162-2020-00046.
IT HAS BEEN REPORTED THAT DURING THE PLANNED REMOVAL OF A LOCKING PROXIMAL FEMUR PLATE, TWO SCREWS FRACTURED. A DELAY GREATER THAN THIRTY MINUTES WAS NOTED. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THIS MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324278 | 3.5MM LOCKING CORTICAL SCREW WITH T15 HEXALOBE, SIZE 26MM | SCREW, FIXATION, BONE | HWC | ORTHOPEDIATRICS, INC | N/A | 171334-J | 00841132110881 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |