FDA Adverse Event Injury Summary report: N

KIT IMPLANTABLE SLIM TIP LEAD, 90CM

MDR report key: 9858248 · Received March 19, 2020

Report

Report Number
1627487-2020-02918
Event Type
Injury
Date Received
March 19, 2020
Date of Event
February 2, 2020
Report Date
August 3, 2020
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
PMP
UDI-DI
05415067027160
PMA / PMN Number
P150004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: ADDITIONAL INFORMATION FOUND - SECTION H6: METHOD CODE SHOULD HAVE BEEN 4114 RATHER THAN 4117.

Description of Event or Problem · 0

RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-023217 1627487-2020-023218 ADDITIONAL INFORMATION FOUND THE PATIENT HAD THEIR 2-L5 LEADS AND 1 OF THEIR S1 LEADS EXPLANTED AND REPLACED TO ADDRESS THE HIGH IMPEDANCE ISSUE. NOTE: IT IS UNKNOWN WHICH S1 LEAD HAD HIGH IMPEDANCE AND WAS REPLACED, AS SUCH BOTH S1 LEADS ARE BEING REPORTED.

Additional Manufacturer Narrative · 1

 THE RESULTS/METHOD AND CONCLUSION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative · 1

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS.  BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 1

RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-02919. IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. REPROGRAMMING WAS UNABLE TO RESOLVE THE ISSUE. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320869 KIT IMPLANTABLE SLIM TIP LEAD, 90CM DRG SLIM TIP LEAD PMP ST. JUDE MEDICAL - NEUROMODULATION MN10450-90A 6151577 05415067027160

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other MODEL # MN10450-50A| MODEL # MN10450-50A - LEAD (X2)| MODEL # MN10450-50A