FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD-PAK

MDR report key: 9853546 · Received March 19, 2020

Report

Report Number
3004123209-2020-00069
Event Type
Malfunction
Date Received
March 19, 2020
Date of Event
January 29, 2020
Report Date
March 26, 2020
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K014067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORDS FOR THE SAM 300P WERE REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TEST HAD BEEN SUCCESSFULLY COMPLETED. THE SAM 300P PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE (B)(6) 2010. THE RANDOM NATURE OF THE EVENTS STORED IN THE MEMORY AND THE 10-MINUTE TIMEOUTS, WOULD SUGGEST THE DEVICE WAS SWITCHING ON AUTOMATICALLY. EXPERIENCE HAS SHOWN THAT IT IS REASONABLE TO CONCLUDE THAT THESE SYMPTOMS MAY BE ATTRIBUTED TO A MEMBRANE FAILURE. THE RETURNED SAM 300P IS NOW OUTSIDE OF ITS WARRANTY PERIOD AND WILL BE SCRAPPED.

Description of Event or Problem · 0

DEFECT FOUND AFTER INVVESTIGATION. MEMBRANE FAILURE. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE DEVICE HISTORY RECORDS FOR THE SAM 300P WERE REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND Q UALITY CHECKS AND TEST HAD BEEN SUCCESSFULLY COMPLETED. THE SAM 300P PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE (B)(6) 2010. THE RANDOM NATURE OF THE EVENTS STORED IN THE MEMORY AND THE 10-MINUTE TIMEOUTS, WOULD SUGGEST THE DEVICE WAS SWITCHING ON AUTOMATICALLY. EXPERIENCE HAS SHOWN THAT IT IS REASONABLE TO CONCLUDE THAT THESE SYMPTOMS MAY BE ATTRIBUTED TO A MEMBRANE FAILURE. THE RETURNED SAM 300P IS NOW OUTSIDE OF ITS WARRANTY PERIOD AND WILL BE SCRAPPED.QAT FROM HEARTSINE TECHNOLOGIES ON THE (B)(6) 2010.

Description of Event or Problem · 1

DEFECT FOUND AFTER INVESTIGATION. MEMBRANE FAILURE. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319968 HEARTSINE SAMARITAN 300P AND PAD-PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD

Patients

Seq Age Sex Outcome Treatment
1