FDA Adverse Event Malfunction Summary report: N

SCR SFTSLK 2.0 PKG.8X20 STER

MDR report key: 9848772 · Received March 18, 2020

Report

Report Number
1219602-2020-00535
Event Type
Malfunction
Date Received
March 18, 2020
Date of Event
February 5, 2020
Report Date
April 10, 2020
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWC
UDI-DI
03596010375391
PMA / PMN Number
K921481
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

ONE 7207259 SOFT SILK 8X20MM SCREW USED IN TREATMENT, WAS NOT RETURNED FOR EVALUATION. DUE TO PRODUCT UNAVAILABILITY, EVALUATION WAS LIMITED. IF FURTHER INFORMATION, BECOMES AVAILABLE, THE COMPLAINT MAY BE REVISITED. PER IFU 1060403: IT IS THE SURGEON¿S RESPONSIBILITY TO BE FAMILIAR WITH THE APPROPRIATE SURGICAL TECHNIQUES PRIOR TO USE OF THIS DEVICE. READ THESE INSTRUCTIONS COMPLETELY PRIOR TO USE.¿ THERE WAS COMPROMISED BONE QUALITY REPORTED. COMPLAINT HISTORY REVIEW INDICATED NO SIMILAR ALLEGATIONS FOR THE LOT NUMBER REPORTED. BATCH REVIEW DID NOT INDICATE A CONDITION, PRODUCT OR PROCEDURE FAILURE THAT SUPPORTED THE ALLEGATION. PRODUCT MET SPECIFICATIONS UPON RELEASE TO DISTRIBUTION. ROOT CAUSE RELATED TO THE MANUFACTURE OF THE DEVICE WAS NOT CONFIRMED. PRODUCT MET SPECIFICATIONS UPON RELEASE TO DISTRIBUTION. NO ADDITIONAL ACTIONS REQUIRED AT THIS TIME.

Additional Manufacturer Narrative · 0

H11: G4: CORRECT AWARE DATE IS (B)(6) 2020.

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING AN ACL REVISION SURGERY, A NEW 8X20MM SOFTSILK SCREW WAS USED, BUT THE TENDON WAS LOOSE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP 8X10MM SOFTSILK SCREW INTO AN ADDITIONAL BONE HOLE. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310368 SCR SFTSLK 2.0 PKG.8X20 STER SCREW, FIXATION, BONE HWC SMITH & NEPHEW, INC. 7207259 50797286 03596010375391

Patients

Seq Age Sex Outcome Treatment
1