FDA Adverse Event Malfunction Summary report: N

ELEVATOR #1

MDR report key: 9848722 · Received March 18, 2020

Report

Report Number
0001032347-2020-00173
Event Type
Malfunction
Date Received
March 18, 2020
Report Date
July 22, 2020
Manufacturer
BIOMET MICROFIXATION
Product Code
EMJ
UDI-DI
00841036028206
PMA / PMN Number
N/A
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE COMPLAINT IS CONFIRMED. THE ELEVATOR #1 (P/N: 09-0259, LOT# 011413A13, QTY 1) WAS RETURNED FOR INVESTIGATION. UPON REVIEW OF THE ELEVATOR, IT WAS NOTED THAT IT HAD A FRACTURE AT THE TIP. THE NON-CONFORMANCE DATABASE WAS REVIEWED FOR THIS PART/LOT# COMBINATION AND THERE WERE NO NON-CONFORMANCES FOUND RELATED TO THIS ISSUE. THERE ARE NO INDICATIONS OF MANUFACTURING DEFECTS. A COMPLAINT REVIEW WAS PERFORMED FOR THIS PART AND LOT NUMBER COMBINATION AND THERE IS AN OCCURRENCE RATE OF 0.22%. THE MOST LIKELY UNDERLYING CAUSE OF THE COMPLAINT IS THAT THE ELEVATOR TIP EXPERIENCED FORCE IN EXCESS OF WHAT IT WAS DESIGNED TO ENCOUNTER. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT (B)(4). IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE RETURNED FOR EVALUATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE INSTRUMENT FRACTURED. ATTEMPTS HAVE BEEN MADE BUT NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313843 ELEVATOR #1 ELEVATOR, SURGICAL, DENTAL EMJ BIOMET MICROFIXATION N/A 011413A13 00841036028206

Patients

Seq Age Sex Outcome Treatment
1