FDA Adverse Event Injury Summary report: N

STEM: QUADRA-C CEMENTED, MIRROR POLISHING STD STEM SIZE 3 12/14

MDR report key: 9848075 · Received March 18, 2020

Report

Report Number
3005180920-2020-00162
Event Type
Injury
Date Received
March 18, 2020
Date of Event
February 18, 2020
Report Date
March 18, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030802249
PMA / PMN Number
K083558
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 28-FEB-2020: LOT 155701: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 7 DECEMBER 2015. EXPIRATION DATE: 2020-11-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR HIP REVISION PERFORMED 3,5 YEARS AFTER HYBRID TOTAL HIP ARTHROPLASTY IN A HEAVY ((B)(6) KG) (B)(6) YEAR OLD PATIENT. NO INFORMATION CONCERNING PATIENT GENERAL HEATH STATUS AND COMORBIDITIES IS AVAILABLE. RADIOGRAPHIC IMAGE PROVIDED SHOWS THE PRESENCE OF SIGNS OF STRESS SHIELDING.

Description of Event or Problem · 1

HIP REVISION AFTER 3 YEARS AND 7 MONTHS FROM THE PRIMARY DUE TO STEM LOOSENING AND INFECTION SUSPICIOUS. ALL IMPLANTS WERE REVISED, NO INFECTION PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311488 STEM: QUADRA-C CEMENTED, MIRROR POLISHING STD STEM SIZE 3 12/14 HIP CEMENTED STEM LZO MEDACTA INTERNATIONAL SA 01.12.043 155701 07630030802249

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention