OAKWORKS MEDICAL CFPM 400 IMAGING TABLE
Report
- Report Number
- 2529571-2020-00004
- Event Type
- Malfunction
- Date Received
- March 17, 2020
- Date of Event
- October 26, 2017
- Report Date
- March 2, 2020
- Manufacturer
- OAKWORKS INC.
- Product Code
- IXR
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
AN INTERNAL PROCESS AUDIT FINDING CREATED A TASK FOR PAR #5 (PREVENTATIVE ACTION RECOMMENDATION) FROM CAPA (B)(4) TO REVIEW THE COMPLAINT DATA AND RETROSPECTIVELY REPORT ANY PREVIOUSLY UNREPORTED COMPLAINTS ASSOCIATED WITH FIELD ACTION RECALL #Z-0703-2018. FOUR COMPLAINTS WERE FOUND DURING THE REVIEW THAT MEET THE CRITERIA WHICH ARE BEING REPORTED.
CUSTOMER REPORTS THEY HAVE A CFM 400 C-ARM TABLE THAT HAS BEEN HAVING AN ISSUE WHERE IT WILL CONTINUE TO GO UP EVEN WHEN NO ONE IS PUSHING THE UP BUTTON, AND USUALLY WON'T STOP UNTIL IT REACHES MAXIMUM HEIGHT. THE ONLY WAY THEY HAVE BEEN ABLE TO STOP IT IS TO USE THE EMERGENCY STOP BUTTON. A REVIEW OF THE FOOT CONTROL PHOTO INDICATES THE CUT WAS ON AN OUTSIDE BUTTON. THIS IMPLIES THE FOOT CONTROL WAS DAMAGED BY CONTACT WITH THE COLUMN.