FDA Adverse Event Malfunction Summary report: N

OAKWORKS MEDICAL CFPM 400 IMAGING TABLE

MDR report key: 9843260 · Received March 17, 2020

Report

Report Number
2529571-2020-00004
Event Type
Malfunction
Date Received
March 17, 2020
Date of Event
October 26, 2017
Report Date
March 2, 2020
Manufacturer
OAKWORKS INC.
Product Code
IXR
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN INTERNAL PROCESS AUDIT FINDING CREATED A TASK FOR PAR #5 (PREVENTATIVE ACTION RECOMMENDATION) FROM CAPA (B)(4) TO REVIEW THE COMPLAINT DATA AND RETROSPECTIVELY REPORT ANY PREVIOUSLY UNREPORTED COMPLAINTS ASSOCIATED WITH FIELD ACTION RECALL #Z-0703-2018. FOUR COMPLAINTS WERE FOUND DURING THE REVIEW THAT MEET THE CRITERIA WHICH ARE BEING REPORTED.

Description of Event or Problem · 1

CUSTOMER REPORTS THEY HAVE A CFM 400 C-ARM TABLE THAT HAS BEEN HAVING AN ISSUE WHERE IT WILL CONTINUE TO GO UP EVEN WHEN NO ONE IS PUSHING THE UP BUTTON, AND USUALLY WON'T STOP UNTIL IT REACHES MAXIMUM HEIGHT. THE ONLY WAY THEY HAVE BEEN ABLE TO STOP IT IS TO USE THE EMERGENCY STOP BUTTON. A REVIEW OF THE FOOT CONTROL PHOTO INDICATES THE CUT WAS ON AN OUTSIDE BUTTON. THIS IMPLIES THE FOOT CONTROL WAS DAMAGED BY CONTACT WITH THE COLUMN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306870 OAKWORKS MEDICAL CFPM 400 IMAGING TABLE IMAGING TABLE IXR OAKWORKS INC. CFPM 400 N/A