FDA Adverse Event
Malfunction
Summary report: N
OAKWORKS MEDICAL CFPM 300 IMAGING TABLE
MDR report key: 9842813
·
Received March 17, 2020
Report
- Report Number
- 2529571-2020-00002
- Event Type
- Malfunction
- Date Received
- March 17, 2020
- Date of Event
- October 12, 2015
- Report Date
- March 2, 2020
- Manufacturer
- OAKWORKS, INC.
- Product Code
- IXR
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
AN INTERNAL PROCESS AUDIT FINDING CREATED A TASK FOR PAR #5 (PREVENTATIVE ACTION RECOMMENDATION) FROM CAPA (B)(4) TO REVIEW THE COMPLAINT DATA AND RETROSPECTIVELY REPORT ANY PREVIOUSLY UNREPORTED COMPLAINTS ASSOCIATED WITH FIELD ACTION RECALL #Z-0703-2018. FOUR COMPLAINTS WERE FOUND DURING THE REVIEW THAT MEET THE CRITERIA WHICH ARE BEING REPORTED.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE UP BUTTON ON THE FOOT CONTROL IS STICKING AND CAUSING THE TABLE TO CONTINUE TO RAISE AFTER THEY RELEASE THE BUTTON. CUSTOMER HAD DAMAGED THE FOOT CONTROL BY LOWERING THE COLUMN ONTO IT. NO PATIENT INVOLVEMENT OR ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 305219 | OAKWORKS MEDICAL CFPM 300 IMAGING TABLE | IMAGING TABLE | IXR | OAKWORKS, INC. | CFPM 300 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |