630G INSULIN PUMP MMT-1715KR 630G
Report
- Report Number
- 3004209178-2020-86656
- Event Type
- Injury
- Date Received
- March 17, 2020
- Date of Event
- March 9, 2020
- Report Date
- October 2, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 00763000173036
- Removal / Correction Number
- Z-0955-2020
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0955-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
UPDATED SUMMARY: THE INSULIN PUMP WENT DEAD AND REPLACED BATTERY IN THE MORNING.
DEVICE WAS RECEIVED WITH A MISSING RETAINER AND MISSING RESERVOIR TUBE O-RING. DEVICE PASSED THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME OR SEATING TEST, BASIC OCCLUSION TEST AND FORCE SENSOR TEST. THE TEST RESERVOIR DOES NOT LOCK IN PLACE AND UNABLE TO PERFORM THE DISPLACEMENT TEST, DISPLACEMENT ACCURACY TEST AND OCCLUSION TEST DUE TO MISSING RETAINER AND MISSING RESERVOIR TUBE O-RING. DEVICE WAS MONITORED WITHOUT THE TEST ENERGIZER BATTERY INSIDE THE BATTERY COMPARTMENT AND REINSERTS IT BACK INTO THE BATTERY COMPARTMENT FOR LESS THAN 10 MINUTES AND NO UNEXPECTED PUMP ERROR 23 ALARM NOTED.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER'S WIFE REPORTED VIA PHONE CALL THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. CUSTOMER STATED THAT INSULIN PUMP RECEIVED POST-RESET RAM CRC ALARM AS WELL AS THE BLACK RING WENT MISSING. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 700 MG/DL AND NOW BLOOD GLUCOSE LEVEL WAS 486 MG/DL WHICH WAS TREATED WITH THE MANUAL INJECTION. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DECLINED FOR HIGH BLOOD GLUCOSE. THE CUSTOMER STATED THAT RESERVOIR ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. CUSTOMER REPORTED THEY WERE ABLE TO CLEAR THE ALARM. CUSTOMER WAS ABLE TO REWIND THE INSULIN PUMP. CUSTOMER STATED THE ALARM OCCURRED IMMEDIATELY AFTER A BATTERY CHANGE. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 304354 | 630G INSULIN PUMP MMT-1715KR 630G | ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715KR | HG32X3Z | 00763000173036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Male | Other | FRN-UNK-RSVR, UNOMED SET.| FRN-UNK-RSVR, UNOMED SET |