FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK SYRINGE BULK STERILE PHARMACY CONVENIENCE PAK

MDR report key: 9837092 · Received March 16, 2020

Report

Report Number
9610847-2020-00101
Event Type
Malfunction
Date Received
March 16, 2020
Date of Event
March 5, 2020
Report Date
March 9, 2020
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: THREE PHOTOS WERE PROVIDED. THEY SHOW A PLUNGER ROD WITH THE RUBBER STOPPER. THE SILICONE APPLIED TO THE STOPPER IS VISIBLE WITH NO RUN OUTS OR ANY NOTICEABLE EXCESS. ONE PHOTO SHOWS A BARREL WITH NO PLUNGER ROD-RUBBER STOPPER. IT SHOWS WHAT APPEAR AN AIR BUBBLE IN THE BARREL WALL OR A DROP OF A SUBSTANCE. FROM THE PHOTO THERE IS NO WAY TO CONFIRM WHAT EXACTLY IS. ROOT CAUSE COULD NOT BE DETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. THIS IS THE 1ST COMPLAINT FOR LOT # 9242252 FOR THIS TYPE OF DEFECT OR SYMPTOM. THERE WAS NO DOCUMENTATION FOR THIS TYPE OF DEFECT DURING THE ENTIRE PRODUCTION RUN OF THESE BATCH #S.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS FOREIGN MATTER (SILICONE) PAST PLUNGER WITH A BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE PAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (2 OF 2 COMPLAINTS). IT WAS REPORTED THAT MULTIPLE SYRINGES WERE OBSERVED TO HAVE WHAT WAS THOUGHT TO BE SILICONE FLUID PAST THE PLUNGER SEAL BUT AFTER REMOVING THE PLUNGER, THE FLUID SOLIDIFIED INTO A CRUST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301664 BD LUER-LOK SYRINGE BULK STERILE PHARMACY CONVENIENCE PAK PISTON SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 9242252

Patients

Seq Age Sex Outcome Treatment
1 Other