FDA Adverse Event Malfunction Summary report: N

PORTEX UNIPERC INNER CANNULAE

MDR report key: 9833561 · Received March 16, 2020

Report

Report Number
3012307300-2020-02018
Event Type
Malfunction
Date Received
March 16, 2020
Date of Event
February 14, 2020
Report Date
July 11, 2020
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
BTO
UDI-DI
15019315107221
PMA / PMN Number
K083031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS COMPLETED ON A SMITHS MEDICAL TRACHEOSTOMY/PVC - PORTEX TUBES PDT UNIPERC . DEVICE RECEIVED 7MM IN PLASTIC BAG WITH PHOTOS PROVIDED.. WHEN INSPECTING DEVCIE IT WAS CONFIRMED THE BLUE LOCKING LEVER IS INCORRECTILY SCREWEED WITH FLANGE BODY WHICH WAS CAUSING THE ISSUE. WE UNSCREWED LOCKING LEVER FROM FLANGE BODY AND ASSEMBLED IT BACK CORRECTLY. WE CAN CONFIRM THAT AFTER CORRECT ASSEMBLY LOCKING MECHANISM WORKS WELL (SEE ATTACHED PHOTO OF SAMPLE AFTER REPAIR). SIMILAR CUSTOMER COMPLAINTS WHERE THE SAME ISSUES RELATED TO UNIPERC FLANGE LOCKING MECHANISM WERE RAISED IN PAST AND THEREFORE PATTERN OBJECT PTN-00039 WAS CREATED.THIS ISSUE WAS ESCALATED TO COMPLAINT TREND PROCESS AS PER DVSOP5007 AND IS TRACKED UNER REVIEW-101279. NO FINDING IN THE DHR WHEN REVIEWING MANUFACTURING.

Additional Manufacturer Narrative · 0

B5 AND H6 WERE UPDATED TO REFLECT NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

INVESTIGATION COMPLETED ON ON A SMITHS MEDICAL TRACHEOSTOMY/PVC - PORTEX TUBES PDT UNIPERC.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED THAT THE BLUE LOCK SYSTEM ON A SMITHS MEDICAL REGARDING A PORTEX UNIPERC INNER CANNULA DID NOT WORK. NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

FOREIGN REPORT SOURCE: (B)(6).

Description of Event or Problem · 1

INFORMATION WAS RECEIVED THAT THE BLUE LOCKSYSTEM ON A SMITHS MEDICAL REGARDING A PORTEX UNIPERC INNER CANNULA DID NOT WORK. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298429 PORTEX UNIPERC INNER CANNULAE TUBE, TRACHEOSTOMY (W/WO CONNECTOR) PRODUCT CODE: BTO BTO SMITHS MEDICAL ASD, INC. 100/899/070CZ 3827426 15019315107221

Patients

Seq Age Sex Outcome Treatment
1