FDA Adverse Event Death Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER

MDR report key: 9825552 · Received March 12, 2020

Report

Report Number
3007042319-2020-01751
Event Type
Death
Date Received
March 12, 2020
Date of Event
March 6, 2020
Report Date
May 21, 2020
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT IS BEING SUBMITTED WITH ADDITIONAL INFORMATION. EVENT DESCRIPTION, INITIAL REPORTER EMAIL, AND H10 UPDATED. ADDITIONAL PRODUCTS: D1: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ DRIVELINE D4: MODEL #: 1103, CATALOG #: 1103, EXPIRATION DATE: 31-AUG-2020, SERIAL OR LOT#: (B)(6), UDI #: (B)(4). D10: NO. H4: MFG DATE: 29-AUG-2018. H5: YES. H6: PATIENT CODE(S): (B)(6) H6: DEVICE CODE(S): (B)(6) H6: FDA RESULTS CODE(S): 3233 H6: FDA CONCLUSION CODE(S): 11 F1 USER FACILITY F2 UF/IMPORTER REPORT NUMBER: MW5093770. F3 USER FACILITY NAME/ADDRESS (B)(6) F4 CONTACT PERSON: (B)(6). F5 PHONE NUMBER: (B)(6). F6 DATE USER FACILITY BECAME AWARE OF EVENT: UNKNOWN. F7 TYPE OF REPORT: INITIAL. F8 DATE OF THIS REPORT: 12-MAR-2020. F11 REPORT SENT TO FDA: YES, DATE SENT: 16-MAR-2020. F12 LOCATION WHERE EVENT OCCURRED: HOME. F13 REPORT SENT TO MANUFACTURER: UNKNOWN. F14 MANUFACTURER NAME AND ADDRESS MFR. NAME: HEARTWARE, INC. ADDL: 14400 NW 60TH AVE CITY: MIAMI LAKES STATE: FL ZIP: 33014. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR INVESTIGATION COMPLETION. PRODUCT EVENT SUMMARY: THE CONTROLLER WAS NOT RETURNED FOR EVALUATION. THE REPORTED LOSS OF POWER EVENT WAS NOT CONFIRMED VIA REVIEW OF THE CONTROLLER LOG FILES SINCE LOG FILES COVERING THE REPORTED EVENT DATE WERE NOT AVAILABLE FOR ANALYSIS. BASED ON THE AVAILABLE INFORMATION, THE MOST LIKELY ROOT CAUSE OF THE LOSS OF POWER CAN BE ATTRIBUTED TO THE REPORTED DISCONNECTION OF BOTH POWER SOURCES FROM THE CONTROLLER, AS DESCRIBED IN THE EVENT DETAILS. INVESTIGATION OF THIS EVENT IS COMPLETED AND THE FILE WILL BE CLOSED. IF NEW INFORMATION IS RECEIVED, THE FILE WILL BE RE-OPENED AND A SUPPLEMENTAL WILL BE SUBMITTED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR INVESTIGATION COMPLETION. PRODUCT EVENT SUMMARY: THE PUMP AND CONTROLLER WERE NOT RETURNED FOR EVALUATION. THE REPORTED "DRIVELINE DISCONNECTED" EVENT WAS NOT CONFIRMED VIA REVIEW OF THE CONTROLLER LOG FILES SINCE LOG FILES COVERING THE REPORTED EVENT DATE WERE NOT AVAILABLE FOR ANALYSIS. OF NOTE, IT WAS REPORTED THAT THE PATIENT ACCIDENTALLY DISCONNECTED THE DRIVELINE. BASED ON THE AVAILABLE INFORMATION, THE MOST LIKELY ROOT CAUSE OF THE REPORTED "DRIVELINE DISCONNECTION" EVENT CAN BE ATTRIBUTED TO THE REPORTED DISCONNECTION OF THE DRIVELINE FROM THE CONTROLLER, AS DESCRIBED IN THE EVENT DETAILS. ADDITIONAL PRODUCTS: D4: SERIAL #: (B)(6). H6: FDA METHOD CODE(S): 4114. H6: FDA RESULTS CODE(S): 3221. H6: FDA CONCLUSION CODE(S): 67. INVESTIGATION OF THIS EVENT IS COMPLETED AND THE FILE WILL BE CLOSED. IF NEW INFORMATION IS RECEIVED, THE FILE WILL BE RE-OPENED AND A SUPPLEMENTAL WILL BE SUBMITTED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

IT WAS FURTHER REPORTED THE PATIENT ACCIDENTALLY DISCONNECTED THE DRIVELINE WHICH LEAD TO THE DEATH AND IT WAS NOT DUE TO A DOUBLE DISCONNECT OF THE POWER SOURCES ON THE CONTROLLER.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANS PLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. CONCOMITANT MEDICAL PRODUCTS: 1103 VAD, IMPLANTED: (B)(6) 2018; DTMA1D1 ICD, IMPLANTED: (B)(6) 2019. INVESTIGATION OF THIS EVENT IS PENDING AND A SUPPLEMENTAL REPORT WILL BE SENT UPON ITS COMPLETION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A LOSS OF POWER FROM A DOUBLE DISCONNECT OF BOTH POWER SOURCES ON THE CONTROLLER. THE PATIENT¿S SPOUSE WAS UNABLE TO RECONNECT THE SOURCES IN A TIMELY MANNER AND THE PATIENT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
286236 HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER VENTRICULAR (ASSIST) BYPASS DSQ HEARTWARE, INC. 1420

Patients

Seq Age Sex Outcome Treatment
1 75 YR Death 694765 LEAD, 419688 LEAD, 5076-52 LEAD.| 694765 LEAD, 419688 LEAD, 5076-52 LEAD