LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER
Report
- Report Number
- 1527736-1997-01265
- Event Type
- Malfunction
- Date Received
- June 12, 1997
- Date of Event
- March 26, 1997
- Report Date
- June 11, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GDO
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- UNKNOWN
Narratives
THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #973508. VISUAL INSPECTIONS & RESULTS: CLIP IN JAW, NO; DAMAGED JAWS, NO; DAMAGED CUTTER, N/A; DAMAGED FEED BAR, NO; DAMAGED FLOOR, NO; DAMAGED HANDLE SHROUDS, NO; DAMAGED OTHER, NO; DAMAGED TIP SHROUDS, NO; DAMAGED TRIGGER, NO; DAMAGED TUBE, NO AND DAMAGED WELD SEAMS, NO. FUNCTIONAL TESTS & RESULTS: ANTIBACKUP FUNCTIONAL, YES; FIRING: FEED CONFORM, YES; FIRING: FORM CONFORM YES; JAWS: HOLD CLIP, YES; JAWS: INSIDE WIDTH AT TIPS, .175; LOCKOUT FUNCTIONAL, YES AND NUMBER OF CLIPS FED AND FORMED, 13. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE INSTRUMENT CYCLED, FED AND FORMED THE REMAINING CLIPS AS DESIGNED. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE THE CLIPS FEED AND FORM PROPERLY. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.
THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE. THE DEVICE WAS SPITTING CLIPS. THE CLIPS DID NOT FALL INTO THE PT. A NEW DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER | ENDOSCOPIC CLIP APPLIER | GDO | ETHICON ENDO-SURGERY, INC. | NA | J45N6R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |