OLYMPIC VAC-PAC
Report
- Report Number
- 3018859-2008-00003
- Event Type
- Other
- Date Received
- January 17, 2008
- Date of Event
- December 11, 2007
- Report Date
- January 17, 2008
- Manufacturer
- OLYMPIC MEDICAL, A DIVISION OF NATUS MEDICAL
- Product Code
- CCX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- RISK MANAGER
Narratives
EVALUATION SUMMARY: THE VAC-PAC WAS RETURNED TO OLYMPIC MEDICAL FOR INVESTIGATION. A VISUAL EVALUATION DETERMINED THAT THE SKIN OF THE VAC-PAC HAD BEEN COMPROMISED BY A SHARP OBJECT (CONSITENT WITH A TOWEL CLIP). NOTE THAT THE VAC-PAC INSTRUCTION MANUAL INCLUDES THE FOLLOWING CAUTIONS UNDER "GENERAL USE": THIS PRODUCT IS FOR USE AS A POSITIONING AID ONLY; DO NOT USE AS AN IMMOBILIZER OR RESTRAINT. AVOID PUNCTURING THE VAC-PAC WITH TOWEL CLIPS OR OTHER SHARP INSTRUMENTS.
CUSTOMER REPORTED, THAT THE OLYMPIC VAC-PAC (A SURGICAL POSITIONING DEVICE) SOFTENED, WHILE BEING USED ON A PT DURING SURGERY. THE DEVICE'S PLUG WAS IN PLACE AT THE TIME THAT IT SUDDENLY SOFTENED (SEVERAL HOURS INTO THE SURGERY). PT FELL FROM OPERATING TABLE AND WAS KEPT OVERNIGHT IN ICU FOR EVALUATION. PT SUSTAINED SOME BRUISING BUT NO OTHER APPARENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPIC VAC-PAC | SURGICAL POSITIONER (CCX) | CCX | OLYMPIC MEDICAL, A DIVISION OF NATUS MEDICAL | SIZE #32 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Hospitalization |