FDA Adverse Event Other Summary report: N

OLYMPIC VAC-PAC

MDR report key: 981815 · Received January 17, 2008

Report

Report Number
3018859-2008-00003
Event Type
Other
Date Received
January 17, 2008
Date of Event
December 11, 2007
Report Date
January 17, 2008
Manufacturer
OLYMPIC MEDICAL, A DIVISION OF NATUS MEDICAL
Product Code
CCX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE VAC-PAC WAS RETURNED TO OLYMPIC MEDICAL FOR INVESTIGATION. A VISUAL EVALUATION DETERMINED THAT THE SKIN OF THE VAC-PAC HAD BEEN COMPROMISED BY A SHARP OBJECT (CONSITENT WITH A TOWEL CLIP). NOTE THAT THE VAC-PAC INSTRUCTION MANUAL INCLUDES THE FOLLOWING CAUTIONS UNDER "GENERAL USE": THIS PRODUCT IS FOR USE AS A POSITIONING AID ONLY; DO NOT USE AS AN IMMOBILIZER OR RESTRAINT. AVOID PUNCTURING THE VAC-PAC WITH TOWEL CLIPS OR OTHER SHARP INSTRUMENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED, THAT THE OLYMPIC VAC-PAC (A SURGICAL POSITIONING DEVICE) SOFTENED, WHILE BEING USED ON A PT DURING SURGERY. THE DEVICE'S PLUG WAS IN PLACE AT THE TIME THAT IT SUDDENLY SOFTENED (SEVERAL HOURS INTO THE SURGERY). PT FELL FROM OPERATING TABLE AND WAS KEPT OVERNIGHT IN ICU FOR EVALUATION. PT SUSTAINED SOME BRUISING BUT NO OTHER APPARENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLYMPIC VAC-PAC SURGICAL POSITIONER (CCX) CCX OLYMPIC MEDICAL, A DIVISION OF NATUS MEDICAL SIZE #32 NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization